Performing a load
z
Ensure that the loading compartment door is closed before loading. Loading
cannot be started when the loading compartment door is open.
z
Do not open the loading compartment door during a loading so as to
prevent adverse effects on loading results and damage to the device.
z
Do not bump, move, vibrate or impact the device during a loading because
it might affect the results.
z
Do not place other instruments such as a centrifuge or vortex on the same
bench where the loader is placed, because other instruments might cause
vibrations to the loader.
z
Pay special attention to the LED status bar, icons, and prompts. If errors
occur, a message appears on the screen. Follow the prompt to troubleshoot
and solve the problem. For information about the troubleshooting, refer
to
Troubleshooting on Page 41 . If the problem persists, contact the
technical support.
In the loading consumables interface, tap
Start
and a confirmation
dialog box appears. Select
Yes
to start loading.
The remaining time for loading appears in the interface and the
following table describes the function of each item in the interface:
Item
Description
Tap to view the loading information, such as DNB ID, and tap
Back
to return to the loading interface.
The status of the device is displayed on the right of this icon.
Tap it and a confirmation message appears. Select
Yes
to pause
loading. Tap it again to resume loading.
Tap it and a confirmation message appears. Select
Yes
to stop
loading.
Summary of Contents for MGIDL-T7RS
Page 7: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...
Page 17: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...
Page 30: ...23 Getting started This chapter describes laboratory preparations and loading preparations ...
Page 55: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...
Page 57: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...
Page 59: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...