Electrical saf
ety
3
z
Do not move the device after the technical support have installed
and debugged the device. Incorrect operation will result in inaccurate
experiment result. If you require to re-position the device, contact the
technical support.
z
Only trained professionals such as doctors, technicians, or laboratory
assistants can operate the device.
z
Do not operate the device in the presence of flammable or explosive
liquids, vapors or gases. Otherwise, it might result in device malfunction.
z
Do not disconnect the power cord under the power-on status.
z
Do not place tubes or plate on the device. Liquids seeping into the device
might cause malfunction.
z
Do not reuse the disposable items.
z
Only the peripheral devices and consumables specified by the manufacturer
can be used.
z
If malfunctions related to fluidics lines (for example, bubble) occur during
an experiment, solve the problems before you restart an experiment.
z
If you have maintenance questions that are not mentioned in this manual,
consult the technical support.
z
Only the components provided by the manufacturer can be used for
device maintenance. Unapproved components might damage the device or
degrade performance.
z
The device is verified before delivery. If serious deviation occurs during use,
contact the technical support for calibration.
Electrical safety
z
Before initial use of the device, assess the electromagnetic environment in
which the device will be used.
z
Do not use the device in close proximity to the sources of strong
electromagnetic fields, such as unshielded sources of radiated emissions.
Radiated signals can reduce the accuracy of the results.
z
Ensure that the device is properly grounded, and the grounding resistance is
less than 4
Ω
. Failure to do so might result in inaccurate experiment results
or even electric shock.
z
Do not remove the device cover and expose the inner component outside.
Otherwise, it might cause electric shock.
Summary of Contents for MGIDL-T7RS
Page 7: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...
Page 17: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...
Page 30: ...23 Getting started This chapter describes laboratory preparations and loading preparations ...
Page 55: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...
Page 57: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...
Page 59: ... T h i s p a g e i s i n t e n t i o n a l l y l e f t b l a n k ...