Site requirements
z
Ensure that the laboratory floor is level and the laboratory table has a
weight capacity of over 200 kg (441 lb).
z
Ensure that the laboratory is free of dust, corrosive and flammable gas, and
heat and wind sources.
z
Ensure that the laboratory is away from direct sunlight and well ventilated.
We recommend that you refer to the standard of a biosafety level (BSL) 2
laboratory.
z
The device should be used with the ultra-pure water machine and
refrigerator, and can be used with the genetic sequencer. Therefore,
enough space should be considered when placing the above instruments.
z
Ensure that enough space is provided around the device for ventilation,
cable connection, and power switch operation.
The following figure indicates distances that are required for optimal
operation and access.
≥
200 mm
(8 inches)
≥
200 mm
(8 inches)
≥
200 mm
(8 inches)
Device
Operation area
Figure 8 Space requirements
Network requirements
Operating environments of the computer
Minimum configuration
Processor: Intel Celeron CPU J1900