Remove the battery door by gently pressing down on the battery cover 
flange, then tip the door out and away from the unit. The O2D2 unit uses 
3 standard quality AA DURACELL ULTRA alkaline batteries, or equivalent. 
Insert the batteries as shown (they will be a tight fit), then replace the door 
by setting the bottom of the door in place and tipping the top in until it 
snaps in place. Take extra care when removing and replacing the batteries 
to not damage the batteries and/or connectors.  

INSERTING OR REMOVING THE TUBING

To INSERT the tubing

, push tubing in until resistance

is felt, then push in a little harder, about another 1/8 inch.
Then give it a gentle tug to make sure it has seated properly.

To REMOVE the tubing

, push tubing in slightly, 

then push in the connector collar while you pull gently 
on the tubing to remove it. 

WHEN REMOVING TUBING;

 

DO NOT

 PULL 

ON THE TUBING WITHOUT PUSHING IN THE 
COLLAR; IT WILL DAMAGE THE CONNECTOR.

When not being used,

 the O2D2 unit, oxygen tubes, cannulas, etc., should be disconnected from the oxygen supply and 

stored in a secure manner to ensure that dirt and debris do not become lodged in the inlet and outlet tubes.  The supplied 
tote bags or a zip-top plastic bag is a good storage container. 

 If the unit is not to be used for 30 days or more, 

remove the batteries.  When using the unit for the first time after storage, replace with new batteries 
to ensure proper operation.  A set of fresh spare batteries should be part of your pre-flight inventory.

Regardless of use replace batteries annually.

 

DO NOT STORE THE EDS UNIT WHILE THE INLET IS UNDER PRESSURE.

 Remove all sources of oxygen pressure and 

secure the unit to ensure it will not become damaged. If the un-pressurized supply line is left hooked to the system, make sure 
that it is first purged with clean dry air or oxygen before the EDS unit is used. 

If the lines are disconnected, the lines 

must be covered so that debris, dust or dirt can’t get in.

MOUNTING THE MH EDS O2D2-2G

You may mount the O2D2 unit to a suitable place using the supplied piece 
of 3M DUAL LOCK tape. First, cut the piece of DUAL LOCK tape in half, 
lengthwise, creating two long rectangles. Peel the protective backing off 
one of the rectangles to expose the adhesive and apply it to the back of 
the unit above the thumb indent for the battery door. DO NOT COVER ANY 
PART OF THE BATTERY DOOR. When a suitable place to mount the unit 
has been found, peel off the protective adhesive backing and press the 
adhesive side to the chosen mounting area. 

INSERTING OR REMOVING THE BATTERIES

STORING THE MH EDS O2D2-2G

1. 

Push

 in the connecting collar

2. 

Pull

 the tube  

 straight back while    
 

pushing

 in the  

  connecting collar.

D-rings

Battery 

Door

3M Tape

Before

Dual-Lock

Apply 

Dual-Lock

Battery 

Door

D-rings

4

A. 

Push

 in the tube

REMOVING

INSERTING

You may also mount the O2D2 by feeding 3/4”- wide straps 
(not included) through the D-rings molded into the unit.

+

+

+

-

-

-

NOTE:
Batteries should be replaced annually or when voltage is low.

  

LITHIUM BATTERIES ARE NOT RECOMMENDED as the Low 
Battery Voltage Indicator WILL NOT function correctly

Summary of Contents for Pulse-Demand EDS O2D2-2G

Page 1: ...atents Pending 800 468 8185 541 923 4100 Fax 541 923 4141 www MHoxygen com 2244 SE Airport Way Suite 100 Redmond OR 97756 7537 THE EDS O2D2 2GTM IS DESIGNED FOR EASY OPERATION CAREFULLY READ THIS INST...

Page 2: ...hly oxidizing gas and can vigorously accelerate combustion It can provide a catalyst for spontaneous combus tion resulting in personal injury or death if not used properly and with caution DO NOT use...

Page 3: ...he O2D2 with a Mountain High four port regulator FPR locate the oxygen input tube clear tube with a short red tube on one end and insert the Red tube into the Red Oxygen In connector on the O2D2 unit...

Page 4: ...ries should be part of your pre flight inventory Regardless of use replace batteries annually DO NOT STORE THE EDS UNIT WHILE THE INLET IS UNDER PRESSURE Remove all sources of oxygen pressure and secu...

Page 5: ...lthy person using a cannula at the given altitudes your needs may be different To determine whether you are receiving enough oxygen in a particular mode you will need to use a pulse oximeter available...

Page 6: ...LOW FAULT EVENT The LED Red Light will flash on and off along with a Hi Lo audio chime for 2 seconds with every event NOTE The FLOW FAULT is not intended as a low or out of oxygen warning It is only i...

Page 7: ...ously and leak oxygen To use the MH EDS O2D2 2G with a third party regulator or built in oxygen system with a pressure higher than 20 psi it is mandatory that you use the MH EDS IN LINE REGULATOR EDS...

Page 8: ...ower Supply with the MH EDS O2D2 2G as other power supplies may not be compatible External 4 5 VDC Power Connector The MyGoFlightAdapter DualMicro is the top of the line DC charger for 12 28 volt syst...

Page 9: ...2 2G unit against defects in materials and workmanship for two 2 years from date of purchase invoice The warranty is non transferable Should any part of the MH EDS O2D2 2G become defective within the...

Page 10: ...to ensure your system is operating properly and upgraded to the latest specifications Any other parts replaced due to damage or field use would be subject to an additional charge Q How long should I e...

Page 11: ...as the amount of dissolved CO2 reduces in the blood so does the need to respirate Unfortunately this also exacerbates hypoxia as less oxygen is inhaled and admitted into the blood as the body has no r...

Page 12: ...EDS Audio Cable Aircraft Input TWO O2D2 2G UNITS DRIVING ONE INPUT The outputs from one O2D2 2G May NOT be tied to the outputs from another O2D2 2G In the case of installing two O2D2 2G units in an ai...

Page 13: ...r we breathe to compensate for either a deficiency on the part of the individual or a de ficiency in the atmosphere in which we are breathing A person may have a respiratory or circulatory impairment...

Page 14: ...c effects during the period in which oxygen is being used Because of the insidious nature of hypoxia a person already mildly hypoxic is very unlikely to even think of using oxygen equipment either int...

Page 15: ...en Systems Transfillers Face Masks with and without mics AMSKM 2100 02 medium with mike Anti Bacterial Face Mask Wipes 00VEN 0077 00 EZ Breathe Quick Disconnect 19053 0009 00 Cylinders E Z Breathe Can...

Page 16: ...ries are used Remove during long term storage External Power 4 5 VDC external power jack with auto battery disconnect with series diode is provided No over voltage protection is provided External Audi...

Reviews: