background image

5

The O2D2 unit is controlled by the Rotary Mode-Selector Switch 

ROTARY MODE-SELECTOR SWITCH SETTINGS AND MODES OF OPERATION

N MODE: “Night” or “Now”

At this setting the O2D2 will immediately start the standard oxygen flow providing 

pulses of oxygen appropriate for an average healthy person using a cannula.

Flow start:

      All altitudes 

 

Use with:

  Cannula

Flow amount:

  Standard Pulse   

Altitude Compensating?

: Yes

The 

D5

 setting will cause the O2D2 unit to delay oxygen flow until it senses a pressure 

altitude of 5,000 ft. and above. The 

D10

 setting  delays oxygen flow until 10,000 ft. and 

above. NOTE: When the barometric pressure is low, it will start operation 

at a slightly 

lower flight altitude

 than when the barometric pressure is high. 

Flow start:

 D5--5,000 ft., D10--10,000 ft. 

Use with:

 Cannula

Flow amount:

  Standard Pulse 

Altitude Compensating?:

 Yes

  D MODES: “Day” or “Delayed” 

Mode-Selector 

Switch Settings

 

  F MODES: “Fat” or “Face Mask” 

The O2D2 has three main modes of user controlled operation:

 

1. Semi-Automatic (N Mode) 

 

2. Fully-Automatic (D5, D10)

 

3. Semi-Automatic (F Modes)

 

NOTE:

 The N and D modes are designed to provide the amount of oxygen needed by an 

average size and healthy person using a cannula at the given altitudes; your needs may 

be different. To determine whether you are receiving enough oxygen in a particular mode 

you will need to use a pulse oximeter (available from Mountain High) to determine your 

blood oxygen saturation (goal is 90-100%) at any given altitude. The selected O2D2 mode 

applies to both users and should be set to accommodate the user with the highest oxygen 

need.  In all modes, the O2D2 provides a pulse of oxygen which increases with altitude, 

i.e., it is altitude compensating.

Mode-Selector as seen 

in the ”OFF” setting

5

F settings 1 to 2 are for small Alps face masks. 

Settings 2 to 3 are for medium sized Alps face masks. 

3 to 4 are for large Alps face masks.

 

F1=

 F2=

 F3= 

 F4=

Small Mask
Medium Mask
Large Mask

NOTE: The F settings are used with the Face Mask or when 

requiring an increased oxygen flow with the cannula.

The F-Mode settings augment the amount of oxygen needed to compensate for the ad-

ditional dead-space plenum associated with face masks. They also can be used with the 

MH cannula in situations where more oxygen may be required over the normal ‘N’ and ‘D’ 

settings. Since F-Mode settings are calibrated for use with the MH Alps face masks, only 
use the approved MH EDS or ALPS face mask.

Flow start:

   All altitudes 

Use with:

  Cannula or face mask

Flow amount:

  Enriched: 

Altitude Compensating:

 Yes

Summary of Contents for Pulse-Demand EDS O2D2-2G

Page 1: ...atents Pending 800 468 8185 541 923 4100 Fax 541 923 4141 www MHoxygen com 2244 SE Airport Way Suite 100 Redmond OR 97756 7537 THE EDS O2D2 2GTM IS DESIGNED FOR EASY OPERATION CAREFULLY READ THIS INST...

Page 2: ...hly oxidizing gas and can vigorously accelerate combustion It can provide a catalyst for spontaneous combus tion resulting in personal injury or death if not used properly and with caution DO NOT use...

Page 3: ...he O2D2 with a Mountain High four port regulator FPR locate the oxygen input tube clear tube with a short red tube on one end and insert the Red tube into the Red Oxygen In connector on the O2D2 unit...

Page 4: ...ries should be part of your pre flight inventory Regardless of use replace batteries annually DO NOT STORE THE EDS UNIT WHILE THE INLET IS UNDER PRESSURE Remove all sources of oxygen pressure and secu...

Page 5: ...lthy person using a cannula at the given altitudes your needs may be different To determine whether you are receiving enough oxygen in a particular mode you will need to use a pulse oximeter available...

Page 6: ...LOW FAULT EVENT The LED Red Light will flash on and off along with a Hi Lo audio chime for 2 seconds with every event NOTE The FLOW FAULT is not intended as a low or out of oxygen warning It is only i...

Page 7: ...ously and leak oxygen To use the MH EDS O2D2 2G with a third party regulator or built in oxygen system with a pressure higher than 20 psi it is mandatory that you use the MH EDS IN LINE REGULATOR EDS...

Page 8: ...ower Supply with the MH EDS O2D2 2G as other power supplies may not be compatible External 4 5 VDC Power Connector The MyGoFlightAdapter DualMicro is the top of the line DC charger for 12 28 volt syst...

Page 9: ...2 2G unit against defects in materials and workmanship for two 2 years from date of purchase invoice The warranty is non transferable Should any part of the MH EDS O2D2 2G become defective within the...

Page 10: ...to ensure your system is operating properly and upgraded to the latest specifications Any other parts replaced due to damage or field use would be subject to an additional charge Q How long should I e...

Page 11: ...as the amount of dissolved CO2 reduces in the blood so does the need to respirate Unfortunately this also exacerbates hypoxia as less oxygen is inhaled and admitted into the blood as the body has no r...

Page 12: ...EDS Audio Cable Aircraft Input TWO O2D2 2G UNITS DRIVING ONE INPUT The outputs from one O2D2 2G May NOT be tied to the outputs from another O2D2 2G In the case of installing two O2D2 2G units in an ai...

Page 13: ...r we breathe to compensate for either a deficiency on the part of the individual or a de ficiency in the atmosphere in which we are breathing A person may have a respiratory or circulatory impairment...

Page 14: ...c effects during the period in which oxygen is being used Because of the insidious nature of hypoxia a person already mildly hypoxic is very unlikely to even think of using oxygen equipment either int...

Page 15: ...en Systems Transfillers Face Masks with and without mics AMSKM 2100 02 medium with mike Anti Bacterial Face Mask Wipes 00VEN 0077 00 EZ Breathe Quick Disconnect 19053 0009 00 Cylinders E Z Breathe Can...

Page 16: ...ries are used Remove during long term storage External Power 4 5 VDC external power jack with auto battery disconnect with series diode is provided No over voltage protection is provided External Audi...

Reviews: