4 - BATTERY INSTALLATION

1.  Put the two AAA batteries into battery compartment in correct polarities.

2.  Push the battery cover horizontally along the arrow shown in figure 3.

WARNINGS:

•  Battery polarities should be correctly installed,  

otherwise, damage may be caused to the equipment.

•  Please remove the batteries if the oximeter will be stored  

for more than 30 days.

•  Please remove the batteries if you want turn off the oximeter.  

Otherwise it is always in sleep mode.

•  Battery may leak or explode if used or disposed off improperly.

5 - OPERATING GUIDE

5.1 Application Method

1.  Remove the battery cover, and insert the two AAA batteries following polarity markings indicated 

inside of the battery compartment, then reposition the cover.

2.  Hold the oximeter with the display facing toward you, slide your finger into the opening probe of the 

device as shown below (Figure 4), until the fingertip touches the built-in stop guide. For best results, 

make sure the finger is centered within the finger guide.

3.  Press the button to activate the oximeter from sleep mode. The measurement interface will appear  

in 3 seconds.

4.  The measurement result will appear on the screen within 10 seconds.
5.  The oximeter will turn to sleep mode automatically within 8 seconds after the finger is removed.

1.

 Press the bottom to open the probe

  

2. 

Insert the finger into the bottom of machine

  

3. 

Press the operating button to activate the oximeter  

Figure 4. Operation Guide

5.2 Notes for Operation

•  The index finger, middle finger, and ring finger are suitable positions for the monitor.

•  Excessive or rapid movement may affect measurement.

•  Improper sensor placement may affect the measurement accuracy.

•  The oximeter can be reused after cleaning and disinfection.

•  The measurement is most accurate when the oximeter and the heart are at the same level.

•  (Only for Clearwave II) The plethysmogram can be used as pulse intensity indicator. 

  The displayed parameters might be unreliable with the disorderly plethysmogram.

•  (Only for Clearwave) The bar graph can be used as pulse intensity indicator. 

  The displayed parameters might be unreliable with un-periodic bar change.

•  The displayed parameters will show invalid indicator as ‘---’ if signal quality is very low.

•  The displayed parameters will show invalid indicator as ’---’ if oximeter fault occurs.
•  The oximeter will shut down after 5 minutes if continuous testing.

6 - SPECIFICATIONS

6.1 Classification

  Type of protection against electric shock............................ II (Internally powered equipment)

  Degree of protection against electric shock....................... Type BF-Applied part (non-defibrillation proof)

  Operating mode............................................................................ Spot checking

  Degree of protection against hazards of explosion......... Ordinary equipment: Not protected 
  Equipment type............................................................................  Fingertip oximeter

6.2 Measurement Specifications

 

 SpO

2

 declared accuracy

 Range(

σ

*)............................................................................ 70% ~ 99% ± 2 digits

 

 

0% ~ 69%: unspecified

 

Resolution........................................................................... 1%

 

Update  Period.................................................................... 1s

 

Averaging Time................................................................. 8s

  PR declared accuracy

 Range(

σ

*)............................................................................ 25 ~ 250 : ± 3 digits

 

Resolution........................................................................... 1bpm

 

Update  Period.................................................................... 1s

 

Averaging Time................................................................. 8s

6.3 Power Requirements

  Specification of alkaline batteries............................................ Two 1.5V (AAA)
  Operating voltage..........................................................................DC 2.5-3V

6.4 Environmental Specifications

 Temperature

   

Operating............................................................................ +41° to + 104°F / 5° to +40°C

   

Storage/Transportation................................................. -4° to + 140°F / -20° to +60°C 

 Humidity

   

Operating............................................................................ 10 ~ 95%, non-condensing

   

Storage/Transportation................................................. 10 ~ 95%, non-condensing 

  Atmosphere Pressure

   

Operating............................................................................ 70 ~ 106kpa

   

Storage/Transportation................................................. 50 ~ 107.4kpa

Thank you for purchasing the SureLife Pulse Oximeter. This manual contains the instructions necessary 

to operate the product safely and in accordance with its function and intended use. Observance of this 

manual is a prerequisite for proper product performance and correct operation and ensures patient and 

operator safety. Read this manual carefully before using the fingertip pulse oximeter. This product is a 

reusable medical device. The lifespan of the product is 2 years. This device is a prescription use device.

1 - SAFETY

1.1 Contraindications

Do NOT use oximeter in a magnetic resonance (MR or CT) environment.

1.2 Warnings

Keep the oximeter away from young children. Small parts such as the battery door, battery, and lanyard 

may be choking hazards.

1.3 Cautions

•  Do not use the oximeter in the presense of flammable anesthetics. 

•  The oximeter needs to be used according to information provided in this user manual. 

•  The equipment is NOT intended for neonate and infant.

•   Do not use a damaged oximeter as it may affect measurement performance.

•   Do not place the oximeter on the same hand/arm when using a blood pressure monitor or cuff. 

•  Do not use the oximeter for more than 5 minutes without relocating the device to another finger.

•   Do not place the oximeter on edema or fragile tissues.

•   Do not use the oximeter as the only basis for making medical decisions, it is intended only to be used 

as additional information that you can give to your licensed health care professional.

•   Do not use the oximeter in high frequency environments such as near electro-surgical equipment.

•   Do not place the oximeter in liquid.

•   Follow local disposal and recycling laws for the oximeter and its components, including the batteries. 

•  Do not stare at the light (infrared is invisible) emitted from the oximeter as it is harmful to the eyes.

•  For clinical limitations and contraindications, please carefully review the medical literature

•  The equipment is just a clinical diagnosis of auxiliary equipment. The physiological data displayed on 

the equipment are for reference only and can not be directly used for diagnostic interpretation.

•  Not suitable for the users with arrhythmia/heart failure/Low perfusion (PI<0.3)/finger trembling.

•  Not suitable for the users with large finger size or exceeding pulse oximeter’s finger measurement 

cavity size.

•  Please don’t use the thumb and tail finger to measure.

•  Discomfort or pain may appear if using the equipment ceaselessly, especially for microcirculation 

barrier patients, it recommended that the equipment should not be used on the same finger more 

than 5 minutes.

•   The oximeter is designed to measure the percentage of arterial oxygen saturation of functional 

hemoglobin. Any of the following conditions may reduce the performance of the oximeter:

 

- Flickering or very bright light 

- Moisture in the oximeter

 

- Body weight less than 44lbs (20kgs) 

- Weak pulse quality (low perfusion)

 

- Venous pulsations 

- Low hemoglobin

 

- Cardiogreen and other intra-vascular dyes 

- Carboxyhemoglobin

 

- Methemoglobin 

- Dysfunctional hemoglobin

 

- Artificial nails or fingernail polish 

- The finger is too cold

 

- Patients with abnormal circulation of finger endings caused by copd.

2 - THE BASICS

The pulse oxygen saturation is the percentage of HbO

2

 in the total Hb in the blood, so-called the O

2

 

concentration in the blood. It is an important bio-parameter for respiration. A number of diseases relating 

to the respiratory system may cause the decrease of SpO

2

 in the blood, furthermore, some other causes 

such as the malfunction of the human body’s self-adjustment, damages during surgery, and the injuries 

caused by some medical checkups would also lead to the difficulty of oxygen supply in human body. 

The corresponding symptoms would appear as a consequence, such as vertigo, impotence, vomit etc. 

Some symptoms may be serious enough to cause significant harm. Therefore, prompt information 

concerning the patients SpO

2

 can help doctors discover potential problems, and is of great importance  

in the clinical and medical field.

2.1 Principle

The Principle of the oximeter is as follows: an experience formula of data process is established making 

use of the Lambert Beer law according to Spectrum Absorption Characteristic of Reductive hemoglobin 

(Hb) and oxyhemoglobin (HbO

2

) in glow & near-infrared zones. The operation principle of the device 

is: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse 

Scanning Recording Technology, so that two beams of different wavelengths of light can be focused  

onto the human nail tip through a perspective clamp finger-type sensor. The signal is obtained and 

measured by a photosensitive element. The information acquired will be shown on screen through 

treatment in electronic circuits and microprocessor.

3 - INTRODUCTION

3.1 Intended Use

The Pulse Oximeter is a portable, convenient, non-invasive device, used to monitor arterial  

hemoglobin oxygen saturation (SpO

2

) and pulse rate. The personal application are adult patients  

(weight: >66lbs (30kgs)) and pediatric patients (weight: 44-66lbs (20-30kgs)). We recommend using  

the index finger, middle finger, or ring finger as suitable positions for the monitor. It is intended for  

spot-checking or attended-care monitoring in home health care and medical facilities.

WARNING:

•  This pulse oximeter is intended to be used only by clinical professionals or under their guidance.  

It must only be used by persons who have received adequate training in its use. Anyone unauthorized 

or untrained must not perform any operation on it.

CAUTIONS:

•  This pulse oximeter is intended for use in 

hospitals, clinical institutions, or healthcare communities.

•  The pulse oximeter is NOT designed for a newborn or infant. For adults and children, a finger thickness 

between 8 - 25.4mm is recommended.

NOTES:

•  The probe is the opening in the middle of the equipment in which the finger is inserted.

•  The probe is the Applied Part of the equipment.

3.2 Features

•  The pulse oximeter is small in size, lightweight, and easy to carry.

•  Easy, one-button operation.

•  There are two modes: sleep and measure.

•  Automatic sleep mode after 8 seconds with no signal.

NOTES:

•  Press the power button to activate oximeter (measure mode) from sleep mode.

Function

Clearwave (FS10D)

Clearwave II (FS20D)

Display

LED

OLED

SpO

2

 parameter measurement

Y

Y

Pulse rate parameter measurement

Y

Y

Bar graph display

Y

Y

Battery display

Y

Y

Automatically enters sleep mode

Y

Y

Pulse waveform display

Y

Four direction display

Y

Symbol

Definition

Symbol

Definition

Symbol

Definition

Symbol

Definition

% SpO

2

The Pulse Oxygen 

Saturation (%)

Serial Number

Battery Positive 

Electrode

BF Type Applied Part

PR

Pulse Rate (BPM)

The Device has no 

Alarm System

Battery Cathode 

Electrode

Manufacturer

IPX2

T

he Product is 

Protected Against 

Harmful Effects of 

Dripping Water per 

IBC 60529. 

Date of  

Manufacturer

Caution, Consult 

Accompanying 

Documents

This Item is Compliant 

with Medical Device 

Directive 93/42/EEC

Power indication (some models have no battery capacity indicator, but have low battery indicator)

3.4 Functions

3.3 Front View

3.5 Symbols

USER MANUAL

Figure 1. Oximeter Schematic Diagram

Figure 3. SureLife Oximeter  

Battery Installation

Light Emitting Diode

Finger

Photo Detector

Probe Body

Figure 2. Front View of Clearwave (FS10D)

Figure 2. Front View of Clearwave II (FS20D)

Pulse Bar Graph

Power Button

Power Indication

SpO

2

 Value

Pulse Rate Value

Power Indication

Pulse Bar Graph

Power Button

Heartbeat Mark

Pulse Waveform

SpO

2

 Value

Pulse Rate Value

Pulse Oximeter

99

70

PRbpm

%SpO

2

70

99

%SpO

2

PR bpm

- -

+

+