XZR200 User’s Manual
12
97337 Issue 4, July 2018
OPERATION
3.7
Continuous Data Streaming
On power up, after the initial 60 seconds warm up period, the XZR200 will automatically
begin to output the measured O
2
concentration and sensor Td as both an averaged and
raw value. The sensor Td value is the measure of the partial pressure of oxygen in
the measurement gas. The O
2
concentration (%) is the Td value scaled by the stored
calibration value.
The averaged values give a stable (smooth) output while the raw un-averaged values
allow the user to detect sudden oxygen changes. The averaged value is the measurement
output on both the 4-20mA and 0 to 10 V DC outputs. The user can alter the averaging
setting to suit the application requirements.
NOTE: To stop or restart the data streaming - the command ‘s’ (lower or
upper case) should be sent to the unit. Data streaming automatically ceases
during calibration.