Technical Data
2.4
thermoMETER CSmicro
10
Measurement Specifications [SF Models]
SF15/ SF02
2W
HS
e
Temperature range IR
-20 ... 350 °C
-30 ... 900 °C
-20 ... 150 °C
(scalable via software)
Spectral
range
8
...14
μ
m 8
...
14
μ
m 8
...
14
μ
m
Optical
resolution 15:1/
2:1 15:1
15:1
CF-lens (optional)
0.8 mm@ 10 mm/
0.8 mm@ 10 mm
0.8 mm@ 10 mm
2.5
mm@
23
mm
Accuracy
1)
±1.5 °C or ±1.5 %
2)
±1.5 °C or ±1.0 %
2)
±1 °C or ±1 %
3)
Repeatability
1)
±0.75 °C or ±0.75 %
2)
±0.75 °C or ±0.75 %
2)
±0.3 °C or ±0.3 %
3)
Temperature
resolution
0.2
°C
0.1
°C
0.025
K
3) 4)
Response time
30 ms (90 % signal)
30 ms (90 % signal)
150 ms (90 % signal)
Warm-up
time
10
min
10
min
10
min
Emissivity/
Gain
0.100
...
1.100
(adjustable
via
software)
Transmissivity
0.100
...
1.000
(adjustable
via
software)
Interface (optional)
USB programming interface
Signal processing
Average, peak hold, valley hold (adjustable via software)
1)
at ambient temperature 23
±
5 °C, whichever is greater; Epsilon = 1; response time 1 s
2)
at object temperatures > 0 °C
3)
at object temperatures > 20 °C
4)
at time constants > 0.2 s