Technical Data
2.6
Optical Charts
The following optical charts show the diameter of the measuring spot in dependence on the distance
between measuring object and sensor. The spot size refers to 90 % of the radiation energy.
The distance is always measured from the front edge of the sensor housing/ CF-lens holder/ air purge.
The size of the measuring object and the optical resolution of the infrared thermometer determine the
maximum distance between sensor and measuring object.
In order to prevent measuring errors the object should fill out the field of view of the optics completely.
Consequently, the spot should at all times have at least the same size like the object or should be
smaller than that.
Fig. 2.1: SF15/2W/HS D:S = 15:1
thermoMETER CSmicro
12