Installation
3.1.3
Air Purge Collars [SF15/ SF02/ 2W/ 2WM-2]
The lens must be kept clean at all times from dust, smoke, fumes and other contaminants in order to avoid
reading errors. These effects can be reduced by using an air purge collar. Make sure to use oil-free,
technically clean air, only.
Fig. 3.8: Standard air purge
collar: fits to the mounting
bracket; hose connection:
3x5 mm [TM-AP-CS]/ for
sensors with a
D:S ratio
≥
10:1
Fig. 3.9: Laminar air purge collar – the
sideward air outlet prevents a cooling
down of the object in short distances;
hose connection: 3x5 mm
[TM-APL-CS]
A combination of the
Laminar air purge collar
with the bottom section of
the Mounting fork allows an
adjustment in two axes.
[TM-APL-CS +TM-MG-CS]
thermoMETER CSmicro
21
The needed amount of air (approx. 2...10 l/ min.) depends on the application and the installation
conditions on-site.