Page 50
Outputs and Inputs
thermoMETER CT
8.2.6
Modbus RTU
8.2.6.1 Serial Interface Parameters
Baud rate:
9600 or 19200, set by user (factory default: 9600)
Data bits:
8
Parity: even
Stop bits:
2
Flow control: off
8.2.6.2 Protocol
The protocol is a Modbus RTU protocol.
8.2.6.3 Installation Overview
Insert the Modbus RTU interface on the CT electronic board and power it with 8 - 36 V.
Select the RS422 mode on the controller.
Fig. 20 Modbus RTU interface