Page 70
Operating
thermoMETER CT
9.
Operating
After powering up the supply voltage the sensor starts an initializing routine for some seconds. During this
time the display will show
INIT
. After this procedure the object temperature is shown in the display. The
display backlight color changes according to the alarm settings
9.1
Restoring Factory Setting
To set the thermoMETER CT back to the factory settings, please first press the
and then the
but-
ton and keep both pressed for approximately 3 seconds.
The display will show
RESET
for confirmation.
Display
Modes [Sample]
Adjustment Range
142.3C
Object temperature (after signal processing)
[142.3 °C]
Fixed
127CH
Sensor temperature [127 °C]
Fixed
25CB
Box temperature
Fixed
142CA
Current object temperature
Fixed
MV5
Signal output channel 1 [0 - 5 V]
0 - 20 = 0 - 20 mA/
4 - 20 = 4 - 20 mA/
MV5 = 0 - 5 V/
MV10 = 0 - 10 V/
TCJ = Thermocouple type J/
TCK = Thermocouple type K
E0.970
Emissivity [0.970]
0.100 ... 1.100
T1.000
Transmissivity [1.000]
0.100 ... 1.100
A 0.2
Signal Output Average [0.2 s]
A---- = inactive/ 0.1 ... 999.9 s