12.4
Saving of Parameter Settings
After switching on the CT sensor, the flash mode is active. This means that changed parameter settings will
be saved in the CT-internal Flash-EEPROM and will also be kept after the sensor is switched off.
In case values should be changed quite often or continuously the flash mode can be switched off by using
the following command:
Decimal: 112
HEX: 0x70
Data, answer:
byte 1
Result:
0 – Data will be written into the flash memory
1 – Data will be not written into the flash memory
If the flash mode is deactivated, all settings will only be kept as long as the unit is powered. If the unit is
switched off and powered on again, all previous settings are lost.
The command 0x71 will poll the current status.