Page 82
Appendix | EtherCAT Documentation
optoNCDT 1900 / EtherCAT
Only PDO mappings that have the same oversampling may be selected. If the sensor was integrated via an ESI file, PDO mappings
are already mutually exclusive when being selected, depending on the PLC software. If this is not the case because the PLC develop-
ment software does not support this feature or the sensor has been integrated online without an ESI file, an invalid combination of
PDO mappings will result in an error message and the process data will not be transferred to the EtherCAT master.
In object 0x1C13 is selected which PDOs are transferred. The PDO mapping objects are selected. The selection process takes place
before switching from PreOP to SafeOP mode.
Example 1:
Startup procedure to output distance 1 (01DIST1):
- Distance 1 is output in 0x6007. In order to transfer 0x6007 in the PDO, the PDO mapping object 0x1A10 must be selected in
0x1C13.
0x1A10 is now mapped, the following PDOs will be transmitted:
As PDOs are grouped, 0x6007:01 contains the linearized distance value; 0x6005:01 still contains the unlinearized value and 0x6006:01
contains the intensity.