Page 71
Appendix | Factory Settings
optoNCDT 1900 / EtherCAT
A 2
Factory Settings
Measurement
averaging
Median, 9 values
Measuring
rate
4 kHz
Peak selection
Highest peak
Language
German
Measuring range
100 % FSO: digital 163768
0 % FSO: digital 98232
LED State
normal operation
Key Select
Supply
voltage
red
flashes yellow appr. 8 Hz
flashes yellow appr. 1 Hz
yellow
green
10 s
t
0
t
1
t
3
t
2
t
4
Fig. 29 Flowchart for starting a sensor with factory settings
t
0
:
Supply voltage is applied
t
1
...
t
3
: both LEDs indicate the start sequence (red-yellow-green each for 1 sec.)
t
2
:
Select
button is pushed during the start sequence (
t
1
...
t
3
)
t
4
:
Select
button is released while the
State
LED is flashing yellow
D
t
=
t
4
-
t
2
;
D
t
(key stroke duration) must be at least 10 sec., max. 15 sec.
Reset to factory setting: Press the
Select
button after having switched on the sensor while the two LEDs light up „red - yellow -
green“. Hold the key pressed. After 10 seconds, the Status LED starts flashing quickly. If you release the key while it flashes quickly,
the sensor is reset to factory settings. If you hold the key pressed for longer than 15 seconds, the sensor is not reset to factory set-
tings.
If the
Select
button is kept pressed when switching on the sensor (or with a reset), the sensor switches to the Bootloader mode.