Page 73
Appendix | Switch between Ethernet Setup Mode and EtherCAT
optoNCDT 1900 / EtherCAT
A 4
Switch between Ethernet Setup Mode and EtherCAT
The sensors start in the last stored operating mode. With the Select button, you can set the sensor to the EtherCAT operating mode.
Press and hold the Select button on the sensor before switching on the power supply on the sensor. Release the button again as
soon as the State LED flashes yellow. Press the button again for approx. 10 to 15 seconds until the State LED flashes red.
State
LED
Firmware installation or
Ethernet/EtherCAT switch
Select
key
Supply
voltage
red
red flashing appr. 8 Hz
yellow flashing appr. 1 Hz
yellow green
t
0
t
1
t
3
t
2
10 ... 15 sec
t
4
EtherCAT operation
Fig. 31 Flowchart for starting a sensor in EtherCAT operation mode
After completion of the firmware installation/switch, the sensor reboots at time
t
4
.
t
0
:
Supply voltage is applied
t
1
:
The State LED starts flashing yellow, the Select button can be released
t
2
:
Within 15 sec. (
t
2
-
t
1
) press Select button again and hold for further 10 ... 15 sec. (
t
3
-
t
2
)
t
3
...
t
4
: Switches from Ethernet setup mode to EtherCAT, duration max. 1 min.
t
4
:
Sensor starts in EtherCAT operation mode.