Page 99
Appendix | EtherCAT Documentation
optoNCDT 1900 / EtherCAT
A 5.5
Error Codes for SDO Services
In case of a negative evaluation of an SDO requirement, a corresponding error code is output in “Abort SDO Transfer Protocol“.
Error code hex
Meaning
0503 0000
Toggle-Bit has not changed
0504 0000
SDO protocol timeout expired
0504 0001
Invalid command registered
0504 0005
Not enough memory
0601 0000
Access to object (parameter) not supported
0601 0001
Attempt to write to a “read-only parameter“
0601 0002
Attempt to write to a “read-only parameter“
0602 0000
Object (parameter) is not listed in the object directory
0604 0041
Object (parameter) is not mapped on PDO
0604 0042
Number or length of objects to be transmitted exceeds PDO length.
0604 0047
General internal device incompatibility
0606 0000
Excess denied because of a hardware error
0607 0010
False data type or length of service parameter is incorrect
0607 0012
False data type or length of service parameter is too large.
0607 0013
False data type or length of service parameter is too small.
0609 0011
Sub-index does not exist
0609 0030
Invalid value of parameter (only for write access)
0609 0031
Value of the parameter too large
0609 0032
Value of the parameter too small
0609 0036
Maximum value falls below minimum value
0800 0000
General error
0800 0020
Data can not be transmitted or saved in application