Assembly Instructions
optoNCDT 2300
Proper Use
The optoNCDT 2300 system is designed for use in industrial and laboratory areas.
It is used for measuring displacement, distance, position and elongation for in-process quality con-
trol and dimensional testing
The sensor may only be operated within the limits specified in the technical data, see instruction
manual, Chap. 3.4. Use the sensor in such a way that in case of malfunctions or failure personnel or
machinery are not endangered.
Take additional precautions for safety and damage prevention for safety-related applications.
Warnings
Avoid unnecessary laser radiation to be exposed to the human body. Switch off the sensor for clean-
ing and maintenance, for system maintenance and repair if the sensor is integrated into a system.
Caution - use of controls or adjustments or performance of procedures other than those specified
may cause harm.
Connect the power supply and the display-/output device in accordance with the safety regulations
for electrical equipment. The power supply may not exceed the specified limits.
> Danger of injury. Damage to or destruction of the sensor.
Avoid continuous exposure to fluids on the sensor. Avoid exposure to aggressive materials (washing
agent, penetrating liquids or similar) on the sensor.
> Damage to or destruction of the sensor.
Avoid shock and vibration to the sensor. Protect the sensor cable against damage.
> Damage to or destruction of the sensor , failure of the measuring device.
Laser Class
The optoNCDT 2300 sensors operate with a semiconductor laser with a wavelength of 670 nm (vis-
ible/red ILD 2300-x) resp. 405 nm (visible/blue ILD 2300-xBL).
The following warning labels are attached to the cover (front and/or rear side) of the sensor housing:
LASER RADIATION
Do not stare into beam
Class 2 Laser Product
IEC 60825-1: 2015-07
P
1mW; P
1.2mW; t=0.5...542 s
0
P
≤
≤
μ
F=1.5...50kHz;
=670nm
THIS PRODUCT COMPLIES
WITH FDA REGULATIONS
21CFR 1040.10 AND 1040.11
LASER RADIATION
Do not stare into beam
Class 2 Laser Product
IEC 60825-1: 2015-07
P
1mW; P
1.2mW; t=0.5...542 s
0
P
≤
≤
μ
F=1.5...50kHz;
=405nm
IEC label
Only for USA
IEC label for ILD2300-x BL
Never deliberately look into the laser beam! Consciously close your eyes or
turn away immediately if ever the laser beam should hit your eyes.
Input and Output
Signal
Designation
Sensor
Pin
Comment
Cable PC2300-x/SUB-D
1
15-pol. Sub-D
+ U
b
1
Supply voltage (11 ... 30 VDC)
1
Ground
2
System ground for supply and ground poten-
tial for RS422-level
9
+Laser on/off
3
optocoupler input, potential-free
Laser off: U
E
≤ 0,8 V (Low)
Laser on: 2,8 V ≤ U
E
≤ 30 V (High)
2
- Laser on/off
4
10
Sync-in/out
2
5
Synchronous- respectively trigger signals,
symmetrically, RS422-Pegel, terminating
resistor 120 Ohm switchable, input or output
selected depending on the synchronization
mode
3
/Sync-in/out
2
6
11
RxD-RS422
7
Serial input RS422, symmetrically,
Internally terminated with 120 Ohm
4
/RxD-RS422
8
12
TxD-RS422
9
Serial output RS422, symmetrically
5
/TxD-RS422
10
13
Tx - Ethernet
11
Ethernet output, potential-free
6
/Tx - Ethernet
12
14
Rx - Ethernet
13
Ethernet input, potential-free
7
/Rx - Ethernet
14
15
Screen
Housing
No galvanic connection to ground
Housing
1) Further cables are available optionaly.
2) In trigger operation the input is used for triggering.
Plug connector: ODU MINI-SNAP, 14-pol., series B, dimension 2,
code F, IP 68. Sensor round pin plug, view: Solder-pin side male cable
connector
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Supply Voltage
,
Nominal value: 24 V DC (11 ... 30 V, max. 150 mA).
ILD 2300
1
2
11 ...
30 VDC
Sensor
Pin
PC2300-x/Y
Color
Supply
Use supply voltage for measurement
instruments only. MICRO-EPSILON
recommends using an optional available
power supply unit PS2020 for the sensor.
1
white
+U
B
2
brown
Ground
Laser on
PC2300-x/Y
+U
3
4
brown
2
1
max.
30 V
ca. 5 mA
Type 1
ILD 2300
B
Laser off:
U OUT < 0.8 V
Laser on:
2.8 V < U OUT < 30 V
white
green
yellow
Ground
MICRO-EPSILON MESSTECHNIK
GmbH & Co. KG
Königbacher Str. 15
·
94496 Ortenburg
www.micro-epsilon.com
X9771234-A091078HDR
*X9771234-A09*
i
If pin Pin 3 with +U
B
and Pin 4 are
not connected with ground, the
laser is off.
Proper Environment
- Protection class:
IP 65 (applies only when the sensor cable is plugged in)
- Operating temperature:
0 °C ... 50 °C (+32 up to +104 °F)
- Storage temperature:
-20 °C ... 70 °C
(-4 up to +158 °F)
-
Humidity:
5 - 95 % (no condensation)
- Ambient pressure:
Atmospheric pressure
Sensor Mounting, Diffuse Reflection
The optoNCDT 2300 sensor is an optical system for measurements with micrometer accuracy.
i
Make sure it is handled carefully when installing and operating!
Mount the sensor only to the existing holes on a flat surface. Clamps of any kind are not
permitted.
Mount the sensor by means of 3 screws type M4. The bearing surfaces surrounding the
fastening holes (through-holes) are slightly raised.
MR SMR + 0.5 MR alpha
2
25
20.5 °
5
26.5
20.5 °
10
35
17.5 °
20
50
13.8 °
SMR = Start of measuring range
MR = Measuring range
In case of bore holes, blind holes, and
edges in the surface of moving targets
the sensor must be arranged in such a
way that the edges do not obscure the
laser spot.
Direct Reflection, Dimensions and free space,
Measuring ranges 2/10/20 mm
Direct reflection
opto
NCDT
ERR
Power on
RUN
EtherCAT
Ethernet
Laser off
In range
Midrange
Error
LASER RADIA
TION
Do not stare into beam
Class 2 Laser Product
IEC 60825-1: 2015-07
P
1mW; P
1.2mW; t=0.5...542
s
0
P
≤
≤
μ
F=1.5...50kHz;
=670nm
Mounting steps
Switch on the supply voltage
on the sensor.
Watch the “State“ LED on the
top side of the sensor.
Position a shining or mirror-
ing measuring object within
the measuring range.
Move the fit-up aid between
sensor and measuring ob-
ject.
The “State“ LED illuminates yel-
low.
Mount the sensor by means
of 3 screws type M4.
Remove the fit-up aid be-
tween sensor and measuring
object.
33.5
(1.32)
4 (.16)
4
(.16)
10 (.39)
alpha
48
(1.89)
2
MR
(direct reflection)
Mounting holes
3x ø4.5 (dia. .18)
for M4 screws
Limits for free
space
13.5
(.53)
67 (2.6)
89 (3.5)
97 (3.8)
75 (3.0)
SMR + 0.5 MR
4.5
(.18)
Diffuse Reflection, Dimensions and free space,
Measuring ranges 2/5/10/20/50/100 mm
MR = Measuring range
SMR = Start of measuring range
MBM = Midrange
MBE = End of measuring range
67 (2.64)
33.5 (1.32)
4 (.16)
4.5
(.18)
89 (3.50)
97 (3.82)
SMR
75 (2.95)
4
(.16)
Y
10 (.39)
MR
48
(1.89)
2
(.08)
Keep this area free
from other light
sources and/or
their reflections
30 (1.18)
15
(.59)
7
7
(.28)
Laser
13.5
(.53)
3x Mounting
holes
ø
4.5
for M4
screws
Limits for
free space
correct
incorrect
(shadow)
MR
2
5
10 20 50 100
SMR 24
24
30 40 45
70
MMR 25 26.5 35 50 70 120
EMR 26
29
40 60 95 170
Y
1.5 3.5 6.5 10 23 33.5
Diffuse reflection
opto
NCDT
ERR
Power on
RUN
EtherCAT Ethernet
Laser off
In range
Midrange
Error
s
μ
LASER RADIATION
Do not stare into beam
Class 2 Laser Product
IEC 60825-1: 2015-07
1mW; P
1.2mW; t=0.5...542 µs
P
≤
F=1.5...50kHz;
=670nm
P0
≤