background image

Quick Guide

Components

 - Sensor
 - Power supply
 - Laptop / PC + USB/Ethernet a Ethernet cable

  Mount the sensor and connect the components together. 

Ethernet Connection

PC2300-0,5/Y

PC2300-x/SUB-D

Patch cable

NCDT

PS2020

230 VAC

PE

 

N

L

PS2020

Measuring object

opto

ERR

Power on

RUN

EtherCAT Ethernet

Laser off

In range

Midrange

Error

LASER RADIATION

Do not stare into beam

Class 2 Laser Product

IEC 60825-1: 2015-07

1mW; P

1.2mW; t=0.5...542 µs

P

P0 

F=1.5...50kHz; 

=670nm

EtherCAT Connection

PC2300-0,5/Y

PC2300-x/SUB-D

PS2020

230 VAC

PE

 

N

L

PS2020

Measuring object

NCDT

opto

ERR

Power on

RUN

EtherCAT Ethernet

Laser off

In range

Midrange

Error

Run

BECKHOFF 

EK1122

X1

X2

Patch cable

LASER RADIATION

Do not stare into beam

Class 2 Laser Product

IEC 60825-1: 2015-07

P

1mW; P

1.2mW; t=0.5...542 s

0

P

μ

F=1.5...50kHz; 

=670nm

Commissioning

The sensor is delivered ex factory with the IP address 

169.254.168.150. 

You can check the IP address of the sensors, that are connected 
to a PC / network, with the  
SensorFinder.exe. program. You will find this program on the 
provided CD. 

  Now start the program SensorFinder.exe and click on the 

button  

Start scan

.

  Select the designated sensor from the list.

  Click the button 

Start 

browser

 to connect the 

sensor with your default 
browser. 

Alternatively: If DHCP is 
enabled and the DHCP 

server is linked with the DNS 
server, an access is possible 
on „ILD2300_SN01234567“ 
(„01234567“ Serial number of 
your sensor). 

  Start a web browser on your PC. Type 

„ILD2300_Serial number“ in the address bar of 
your web browser.

Interactive websites for programming the sensor 
now appear in the web browser.

  If you have changed any settings, go to the 

menu 

Preferences

 and click on the button 

Save Setup 

to store your settings. 

Select a Measuring Program

  Go to the menu  

Preferences > Measuring program

.

 Select 

Diffuse reflection 

from the measurement arrangement list. Confirm with 

Apply

.

Select a Measuring Rate

  Go to the menu  

Preferences > Measuring rate

.

Start with a medium measuring rate. Select a measuring rate from the list. Confirm with 

Apply

.

Select a Digital Interface

  Go to the menu 

Preferences > Digital interfaces > Selection of digital 

interfaces

.

 Select 

Web Diagram

 from the list. Confirm with 

Apply

.

Store the Settings

  Go to the menu 

Preferences > Load/save settings

.

 Select 

Parameter set

 from the data selection list, a parameter set number and click on the 

button 

Save

.

Position a Measuring Object

  Position a measuring object (target) as possible in the midrange.

100 %

50

0

SMR

SMR                   MMR                            EMR

Displacement

Signal

NCDT

opto

ERR

Power on

RUN

EtherC

AT

Ethernet

Laser off

In range

Midrange

Er

ror

Measuring object

Measuring range

LASER RADIA

TION

Do not stare into beam Class 2 Laser Product

IEC 60

82

5-1: 201

5-0

7

P

1mW

; P

1.2

mW

; t=0.5...5

42

s

0

P

μ

F=1.5...5

0kHz; 

=67

0nm

The Status LED Status on the sensor indicates the position of the measuring object to the sensor.

LED

Color

Labeling

Meaning

Status

off

Laser off

Laser beam is switched off

green In range

Sensor operates, measuring object within measuring range

yellow Midrange

Measuring object is in midrange

red

Error

Measuring object outside measuring range,   
reflection is to low

Displacement Measuring

  Go to the menu 

Measurement

.

  Disable the 

Autoscale 

function and click on the 

Start 

button. 

Thickness Measurement

  Got to the menu 

Preferences > Measuring program 

and select 

Direct reflec-

tion - thickness measurement 

from the measurement arrangement list. 

  Select the target material from the material list. Confirm with 

Apply

.

Store the Settings

  Go to the menu 

Preferences > Load/save settings

.

 Select 

Parameter set

 from the data selection list, a parameter set number and click on 

the button 

Save

.

Read the detailed instruction manual before using the sensor. The manual is available online on
www.micro-epsilon.com/download/manuals/man--optoNCDT-2300--en.pdf or on the supplied 
CD.

Diffuse Reflection, Dimensions for Measuring Range 40, 200 and 300 mm

MR

40

200

300

SMR

175

130

200

MR

195

230

350

EMR

215

330

500

a

22.1 °

25.1 °

18.3 °

j

21.9 °

16.7 °

12.2 °

e

21.8 °

13.1 °

9.6 °

A

101

91.6

99.4 °

B

86

76.0

81

Start of measuring range

End of measuring range

SMR

MR

150 (5.90)

80 (3.15) 70 (2.76)

140 (5.51)

130 (5.52)

15 (.59)

12 (.47)

17.5 (.69)

35 (1.38)

A

B

3 x mounting
hole ø4.5
(.18 dia.)

ø5 (1.20 dia.)

Reviews: