Page 3
scanCONTROL 3010 / 3012
scan
CONTROL
COMPLIES WITH 21 CFR 1040.10 AND 1040.11
EXCEPT FOR CONFORMANCE WITH
IEC 60825-1 ED. 3., AS DESCRIBED IN
LASER NOTICE NO. 56, DATED MAY 8, 2019
LASER RADIATION
DO NOT STARE INTO BEAM
OR EXPOSE USERS OF
TELESCOPE OPTICS
CLASS 2M LASER PRODUCT
IEC 60825-1: 2014
P
10 mW, P
10 mW; E 6 mW/cm²;
0
P
λ
= 658nm; F = 0...25kHz, t = 7µs...∞
LASER
APERTURE
Fig. 5 Sensor with
laser labels
5.
Connections, LED Displays
1 2
1 Multifunction port (power supply, I/O)
2 Ethernet port
Multifunction Port
Designation
Sensor connector
pin
Cable color
PCR3000-x
Notes
+Ub
9
Red
+ 11 V - 30 V DC
(rated value 24 V); max. 5 W
GND
2
Blue
0 V
+Laser on/off
3
White
Available with SI option
-Laser on/off
1
Brown
RS422
12
Red-blue
RS422
input or output
/RS422
11
Gray-pink
GND RS422
10
Purple
Ground connection RS422
In1
4
Green
Switching input In1
In2
6
Yellow
Switching input In2
In3
8
Gray
Switching input In3
In4
5
Pink
Switching input In4
GND-In
7
Black
Ground connection In
Shield
Housing
Black
Not electrically connected to GND