background image

Page 3

scanCONTROL 3010 / 3012

scan

CONTROL

COMPLIES WITH 21 CFR 1040.10 AND 1040.11 

EXCEPT FOR CONFORMANCE WITH 

IEC 60825-1 ED. 3., AS DESCRIBED IN

LASER NOTICE NO. 56, DATED MAY 8, 2019

LASER RADIATION

DO NOT STARE INTO BEAM

OR EXPOSE USERS OF 

TELESCOPE OPTICS

CLASS 2M LASER PRODUCT

IEC 60825-1: 2014

P

10 mW, P

 10 mW; E  6 mW/cm²;

0

P

λ

 = 658nm; F = 0...25kHz, t = 7µs...∞

LASER

APERTURE

Fig. 5 Sensor with 
laser labels

5. 

Connections, LED Displays

          1             2

1 Multifunction port (power supply, I/O) 
2 Ethernet port

Multifunction Port

Designation

Sensor connector 
pin

Cable color 

PCR3000-x

Notes

+Ub

9

Red

+ 11 V - 30 V DC 

(rated value 24 V); max. 5 W

GND

2

Blue

0 V

+Laser on/off

3

White

Available with SI option

-Laser on/off

1

Brown

RS422

12

Red-blue

RS422  

input or output

/RS422

11

Gray-pink

GND RS422

10

Purple

Ground connection RS422

In1

4

Green

Switching input In1

In2

6

Yellow

Switching input In2

In3

8

Gray

Switching input In3

In4

5

Pink

Switching input In4

GND-In

7

Black

Ground connection In

Shield

Housing

Black

Not electrically connected to GND

Summary of Contents for scanCONTROL 30 BL Series

Page 1: ...etergents cooling emulsions Damage to or destruction of the sensor Read the detailed operating instructions before operating the sensor These instructions are available online at www micro epsilon com...

Page 2: ...he sensor housing LASER RADIATION DO NOT STARE INTO BEAM OR EXPOSE USERS OF TELESCOPE OPTICS CLASS 2M LASER PRODUCT IEC 60825 1 2014 P 10 mW P 10 mW E 6 mW cm 0 P 658nm F 0 25kHz t 7 s LASER RADIATION...

Page 3: ...ns LED Displays 1 2 1 Multifunction port power supply I O 2 Ethernet port Multifunction Port Designation Sensor connector pin Cable color PCR3000 x Notes Ub 9 Red 11 V 30 V DC rated value 24 V max 5 W...

Page 4: ...ts of the multifunction port can either be used for encoder input trigger input or for load previously stored user modes The signal levels are switchable for all switching inputs between LLL TTL logic...

Page 5: ...is setting is activated by default and has priority over search the link local net The sensor scanCONTROL 30xx supports Power over Ethernet If the sensor is connected to a POE capable network adapter...

Page 6: ...us RTU protocol Measurement value transmission in ASCII format 7 System Requirements scanCONTROL Configuration Tools The following minimum system requirements are necessary Windows 7 Windows 8 or 8 1...

Page 7: ...Close unused connection plugs with the included protective caps Turn on the power supply 24 VDC Please wait until the scanCONTROL measuring system is recognized by the PC This may take a few seconds N...

Page 8: ...scanCONTROL and the PC On the left side you can adjust the settings for your measurement task the right side shows the profile data and information about the measurement 10 How to Access Profile Data...

Page 9: ...us systems Furthermore all scanCONTROL sensor settings can be set via the 2D 3D Gateway e g Laser on off or switch User Mode Sensor parameterization Power supply Triggering Encoder Synchronization Mod...

Page 10: ...ply OUT 24 V 0 V 0 V System supply In 24 V 0 V OU filter module 2 5 3 0 5 7 LINK ACT MS I O ETHERNET NS 8 7 6 5 4 3 2 1 N O LINK ACT 1 2 X 1 X 2 8 1 ON 24V X3 0V 0 WBM 255 DHCP OU fieldbus coupler L1...

Page 11: ...e Digital output terminals need a field supply of 5 VDC or 24 VDC depending on the module type After mounting the modules install the required wiring Connect the System supply out contacts of the OU F...

Page 12: ...ocumentation the Status and Error Messages and Notes sections in the scanCONTROL Configuration Tools operating instructions You will find details about the individual programs in the respective operat...

Reviews: