1.6
Proper Environment
-
Protection class
1
: IP40 (only applies in the case of connected output connectors and/
or installed protective caps)
-
Temperature Range operation:
+5 ... +40 °C (+41 ... +104 °F); for free air circulation
It is recommended to recalibrate the sensor in case of temperature fluctuations du-
ring operation of more than 5 K.
For temperature differences of more than 10 K (e.g. after transport to another loca-
tion), the sensor may only be put into operation after a period of approx. 1 h.
-
Temperature range storage: -10 … +50 °C (+14 ... +122 °F);
at max. 90 % relative humidity
-
Ambient pressure:
Atmospheric pressure
-
Humidity
Operation:
20 … 80 % (non-condensing)
Storage:
10 … 90 % (non-condensing)
2.
llumination
The surfaceCONTROL 2500 sensor has an LED lighting unit with a high-power LED of
wavelength 462 nm (blue), 528 nm (green) or 612 nm (red) depending on the model.
The LEDs belong to the risk group 2 according to EN 62471: 2010-06.
Do not look into the lens. Consciously close your eyes or immediately turn away if the
optical radiation enters the eye.
> Risk of damage to the eyes by optical radiation!
The following information label is attached to the front of the sensor housing:
Fig. 1 LED warning label
1) The protection class does not apply for the optical sections during operation as their
soiling / contamination results in adversely affecting or failure of the function.
CAUTION