Troubleshooting
5
connection. Check that the cart voltage
changes as you manually lift the mass
(using a hex wrench).
Very bad sound during drop
Sphere signal is not present because of the
cable is not connected or a there is a bad
cable/connector. Look for a sphere signal
that changes as the mass is locked.
Sharp spikes (faster than 1ms) on the sphere
signal during drop with a bad sound during
drop
Bad cable/connector
Many bad drops occur with extremely high
residuals. [High number of drop-outs]
The cart travel is not aligned with the
vertical. The bubble levels on the tripod
need to be reset or the dropping chamber
is not sitting in the tripod tray correctly.
Dropper guide rods are not straight. Can
be due to snubbers tightened
asymmetrically or because of stress from
bad alignment of rods in the top ring.
Table 5-1
Dropping Chamber Problems and Solutions
5.2.2.INTERFEROMETER AND LASER
To achieve the best possible results, the interferometer optics must be clean
and correctly aligned. In some cases, it is possible to detect alignment
problems during setup. The operator should always be extremely observant
while leveling the interferometer, and adjusting beam coincidence and
verticality so any problems in misalignment can be detected and corrected
before observations begin. Table 5-2 gives a list of problems and possible
solutions which are related to the interferometer.
5-4
Problem
Solution
WEO laser not producing beam
Check cable connections. If the front
panel show no sign of power check the
fuse and power switches.
WEO laser power too low
Laser optics need cleaning.
Alignment of back laser mirror should be
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...