Troubleshooting
5
instrument, or optical feedback)
Large residual amplitude
Check parameter file, do not fit past point
170
Bad cables, especially PIO ribbon cable
Bad grounding of fringe signal
10nm residual frequency swept
Wrong laser type selected
Incorrect laser modulation frequency
entered in the fg5param.dat file
No fringes at all
No power to APD.
One/both beam blockers are
pushed in
Very bad alignment of the
test/reference beams. Look into
the telescope and make sure that
the test and reference beams are
overlapped. Also make sure that
the beams are overlapped in the
fringe viewer with the twiddler
Fringe amplitude varies as the cart is
moved up or down.
The iodine-stabilized laser is
running multi-mode. This can
indicate the laser servo gain being
set too high.
Table 5-2
Interferometer Problems and Solutions
5-6
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...