Troubleshooting
5
5.2.4.SYSTEM CONTROLLER
Since the system controller performs real time gravity
computations/corrections, records data, and initiates the dropping sequence,
a failure of this component results in a catastrophic system failure. Table 5-4
gives a list of problems and possible solutions for the system controller. The
primary source of all problems encountered with the controller units have
been with poor cable connections or the computer to docking station
connection and lock in.
Problem
Solution
Olivia locks and the status line displays
reading barometer or getting
environmental.
Turn of the computer and
disconnect and reconnect the serial
cable. Use the GETBAR program
to test the barometer.
Olivia ends with the:
Error Number => 00308
Original Command => OUTPUT 16 …
The UTIC is not connected to the
system or it is turned off.
Olivia ends with the:
Error Number => 00308
Original Command => OUTPUT 14 …
The IOtech ADC 16A is not
connected to the system or is not
turned on.
OLIVIA locks and the UTIC displays
BINARY DATA.
No Fringe signal is hooked up or
one of the beam blockers is
inserted. The condition can be
cleared by manually forcing a drop
using the internal trigger for the
dropping controller.
Gravity value is wrong
Check all gravity corrections
especially vertical transfer to the
specified height. Check
latitude/longitude and date/time.
Check laser peak lock
hardware/software
5-9
Table 5-4
System Controller Problems and Solutions
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...