Troubleshooting
5
5.2.6.ENVIRONMENTAL SENSORS
PACKAGE
The environmental sensors package is an optional system for the FG5.
Therefore, if there is a major failure, it is possible to continue observations.
Occasionally, the operator can obtain temperature and pressure data from an
alternate source. If the pressure data is accurate enough (
±
1 mb or better),
local barometric pressure attraction corrections accurate to better than one
µ
Gal can still be computed and applied during post processing. Possible
problems and solutions pertaining to the environmental sensors package are
shown in Table 5-6.
Problem
Solution
Room temperature reading incorrectly. Change battery in thermocouple
mounted on tripod tray.
Turn on temperature probe
Attach thermocouple to
temperature probe
Attach BNC between temperature
probe and BNC16 channel 1H
Gravity is 300
µ
Gal too high
The system is not reading the
barometer. Make sure the
barometer is hooked up correctly.
Table 5-6
Environmental Sensors Package Problems and Solutions
5-12
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...