Adjustment and Maintenance
4
Figure 4-3
Beam expander
4.2.2.6.2.Collimating lens:
Insert the collimating lens (128mm focal length) into the mount and adjust the
position of the lens to be about 153.4mm from the optical center of the
focusing lens. Look through the telescope at the spot and minimize the
diameter of the spot in the telescope by moving the collimating lens back and
forth. It is important to make sure that the crosshairs are in focus for the
individual performing this alignment. It is critical that the telescope has
already been adjusted for infinity focus using an autocollimation technique
and locked in place.
4.2.2.7.Leveling the Tripod Tray
Now that the beam has been steered through the optical isolator and beam
expander, it remains to use mirrors #3 and #4 to steer the beam vertically off
of beamsplitter #1 through the dropping chamber. This adjustment will be
made assuming that the interferometer is referenced to the bubble levels in
the tripod tray. First, we must verify or set the tripod bubble levels so that
they are aligned with the cart travel inside the dropping chamber.
First re-install the top middle cover on the interferometer base (leave
the two end plates removed for access). Pull the alignment pins out and
make sure that the interferometer base handles are in the upright position.
Put the tripod tray onto the interferometer base without the feet. Place the
dropping chamber onto the tripod tray and clamp it down.
4.2.2.8.Making the test beam vertical using the alcohol
reference surface
4-12
To make the test beam vertical, one must place a dish of alcohol on the floor
below the interferometer base (with the Superspring removed). It is good to
use a dish with enclosed sidewalls to reduce wind vibrations and enough
width (at least 50mm) to avoid a severe meniscus. Looking through the
telescope one can see the reference beam and the return beam from the
alcohol surface. The beam is traveling vertically when these two beams
overlap in the telescope. Level the beam by adjusting the legs of the
interferometer base so that the test and reference beam are coincident in the
telescope. Note that this procedure does not rely on the dropper being
vertical. At this point the bubble levels are not necessarily leveled, but this not
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...