Adjustment and Maintenance
4
a problem since the following procedures will serve to define the correct level
of the tripod tray. Now put the feet under the tripod tray and lift all three
until they just barely make contact with the nylon balls on the tripod legs.
Then lift all three feet one revolution so that the tripod is not making contact
with the interferometer base. You may now level the dropping chamber
using the bubble levels (if you believe they are correct). If the bubble levels
are not believed to be correct, you can ignore them for the moment. In any
case, the test beam is vertical and now we must align the dropping chamber
so that the cart travel is also vertical.
4.2.2.9.Leveling the Dropper
4-13
Make sure that the beam coming back from the dropping chamber is full and
is not clipped or smeared on the edges. If the beam doesn’t go through the
dropper, you may have to adjust mirrors 3 & 4 (see the next section for this
procedure) or you can alternatively translate the dropper. Then switch the
dropper controller to OSC mode in order to make the dropper move
smoothly up and down. Initiate the dropper in OSC mode by hitting RESET
then INIT. The goal is to level the tripod so that the laser beam coming back
from the dropper cornercube does not translate as the cart travels up and
down. The laser spot can be monitored by eye, but is much better if
monitored using a quadrant photodiode such as the translation detector
supplied with the FG5. Put the translation detector on the floor and attach X
and Y outputs to channels 1 and 2 on an oscilloscope. Set the scope on XY,
200 mV/div. Move the translation detector into the laser beam path so that
the spot on the scope is near the origin. Throughout this procedure it is
important that the spot stay near the origin since there are non-linear effects
on the edges of the detector. Note the magnitude of movement of the spot on
the scope as the cart raises and lowers. Adjust the tripod feet until the spot
movement is minimized, moving the detector on the floor to keep the spot
near the origin. Note which direction the beam moves as the cart raises: this
is the direction the cart is moving as it is lifted. Raise the leg which is in the
same direction as the movement of the cart as it moves upwards. When
necessary, increase the sensitivity of the scope and again minimize the spot
movement by adjusting the tripod feet. Repeat this procedure until the spot
movement is below 50 mV in both axes. Make sure the tripod does not touch
the interferometer and that the beam returns from the dropping chamber
without being clipped.
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...