92
Micro Motion
®
Model 2200S Transmitters
Operating the Transmitter
8.7.2
Controlling totalizers and inventories
Table 8-1 shows all of the totalizer and inventory functions and which configuration tools you can use
to control them.
With the display
Table 8-2 shows how you can control the totalizers and inventories with the display.
Table 8-1
Totalizer and inventory control methods
Function name
Display
Communicator
ProLink II
Start/stop all totalizers and inventories
Yes
(1)
(1) If enabled. See Section 5.2.4.
Yes
Yes
Reset mass totalizer value only
Yes
Yes
Yes
Reset volume totalizer only
Yes
Yes
Yes
Simultaneously reset all totalizer values
Yes
Yes
Yes
Simultaneously reset all inventory values
No
No
Yes
(2)
(2) If enabled in the ProLink II Preferences window.
Reset mass inventory value only
No
No
Yes
Reset volume inventory value only
No
No
Yes
Table 8-2
Totalizer and inventory control with the display
To accomplish this
Press this sequence of buttons
Stop all totalizers and
inventories
(1)
(1) This feature may be enabled or disabled. See Section 5.2.4.
•
Scroll
until a totalizer value appears (the word
TOTAL
appears in the lower left corner of
the display). It does not matter whether the total is mass or volume.
• Select.
•
Scroll
until
STOP
appears beneath the current totalizer value.
•
Select
(
YES
alternates with
STOP
).
•
Select
(all totalizers and inventories stop).
•
Scroll
to
EXIT
.
Start all totalizers and
inventories
•
Scroll
until a totalizer value appears (the word
TOTAL
appears in the lower left corner of
the display). It does not matter whether the total is mass or volume.
• Select.
•
Scroll
until
START
appears beneath the current totalizer value.
•
Select
(
YES
alternates with
START
).
•
Select
(all totalizers and inventories start).
•
Scroll
to
EXIT
.
• Select.
Reset mass totalizer
•
Scroll
until the mass totalizer value appears.
• Select.
•
Scroll
until
RESET
appears beneath the current totalizer value.
•
Select
(
YES
alternates with
RESET
).
•
Select
(mass totalizer resets).
•
Scroll
to
EXIT
.
• Select.
Reset volume (liquid or
gas) totalizer
•
Scroll
until the volume totalizer value appears.
• Select.
•
Scroll
until
RESET
appears beneath the current totalizer value.
•
Select
(
YES
alternates with
RESET
).
•
Select
(volume totalizer resets).
•
Scroll
to
EXIT
.
• Select.
Summary of Contents for 2200S
Page 8: ...vi Micro Motion Model 2200S Transmitters ...
Page 12: ...4 Micro Motion Model 2200S Transmitters ...
Page 32: ...24 Micro Motion Model 2200S Transmitters ...
Page 38: ...30 Micro Motion Model 2200S Transmitters ...
Page 68: ...60 Micro Motion Model 2200S Transmitters ...
Page 92: ...84 Micro Motion Model 2200S Transmitters ...
Page 102: ...94 Micro Motion Model 2200S Transmitters ...
Page 110: ...102 Micro Motion Model 2200S Transmitters ...
Page 130: ...122 Micro Motion Model 2200S Transmitters ...
Page 134: ...126 Micro Motion Model 2200S Transmitters ...
Page 150: ...142 Micro Motion Model 2200S Transmitters ...
Page 156: ...148 Micro Motion Model 2200S Transmitters ...
Page 157: ......