2
European installations
This Micro Motion product complies with all applicable European
directives when properly installed in accordance with the instructions in
this quick reference guide. Refer to the EC declaration of conformity for
directives that apply to this product.
The EC declaration of conformity, with all applicable European
directives, and the complete
ATEX Installation Drawings and Instructions
are available on the internet at www.micromotion.com/atex or through
your local Micro Motion support center.
Installation kit
For mounting in a rack, the Model 3300/3500 installation kit includes the
following parts:
•
One DIN standard 41612, Type D connector for input/output wiring,
with solder tails or screw terminals
•
(Model 3500 only) One DIN standard 41612, keyed Type D connector
for sensor wiring, with solder tails or screw terminals
•
One plug-in connector for power supply wiring
•
Four (Model 3300) or six (Model 3500) slotted cheese-head machine
screws, size M2.5x8, for securing wiring connectors to the rack
STEP 1.
Choosing a location
Choose a location for the transmitter based on the requirements described
below.
Environmental requirements
Install the transmitter where ambient temperature is between
–
4 and
+140 °F (–20 and +60 °C).
If multiple applications platforms are installed, provide at least 1 U
(1 HE) of vertical space between racks to ensure proper ventilation. See
Figure 1, page 3.