STEP 4.

Connecting input and output wiring

Connect input and output wiring to the appropriate terminals on the 
input/output wiring connector, which is the far right connector. Refer to 
Table 1 and to the card that is inserted into the sleeve on the top panel 
(shown in Figure 4).

Use 24 to 16 AWG (0,25 to 1,5 mm

2

) twisted-pair shielded wire.

Ground the shields at a single point only.

Figure 4.

Input/output wiring terminal card

STEP 5.

Connecting the Model 3500 to the sensor

To connect the Model 3500 transmitter to a Micro Motion sensor, follow 
the steps below. If you are installing the Model 3300 applications 
peripheral, this step is not required.

Table 1.

Input/output wiring terminals

Terminal number

Designation

c 2+

a 2 –

Primary 4–20 mA output

c 4 +

a 4 –

Secondary 4–20 mA output

c 6 +

a 6 –

Frequency input

c 8 +

a 8 –

Discrete input 1

c 10 +

a 10 –

Discrete input 2

c 12 +

a 12 –

Frequency output

c 14 +

a 14 –

Discrete output 1

c 16 +

a 16 –

Discrete output 2

c 18 +

a 18 –

Discrete output 3

c 32 (B line)

a 32 (A line)

RS-485 output

Summary of Contents for 3500

Page 1: ... 2003 For online technical support use the EXPERT2 system at www expert2 com To speak to a customer service representative call the support center nearest you In the U S A phone 1 800 522 MASS 1 800 522 6277 In Canada and Latin America phone 303 530 8400 In Asia phone 65 6770 8155 In the U K phone 0800 966 180 toll free Outside the U K phone 31 0 318 495 67 ...

Page 2: ... information about hazardous applications refer to Micro Motion ATEX UL or CSA installation instructions shipped with the transmitter or available from the Micro Motion web site WARNING Hazardous voltage can cause severe injury or death Install transmitter and complete all wiring before supplying power CAUTION Improper installation could cause measurement error or flowmeter failure Follow all inst...

Page 3: ...t includes the following parts One DIN standard 41612 Type D connector for input output wiring with solder tails or screw terminals Model 3500 only One DIN standard 41612 keyed Type D connector for sensor wiring with solder tails or screw terminals One plug in connector for power supply wiring Four Model 3300 or six Model 3500 slotted cheese head machine screws size M2 5x8 for securing wiring conn...

Page 4: ...s see Figure 2 page 4 Height 128 mm 3 U or 3 HE Width 142 mm 28 HP or 28 TE Depth 160 mm The Model 3300 3500 conforms to DIN standard 41494 Part 5 IEC 297 3 for 19 inch 486 2 mm racks Up to three enclosures fit into one rack See Figure 1 1 U 1 HE 1 750 inches 44 5 mm 3 U 3 HE 3 U 3 HE 3 U 3 HE 1 U 1 HE 1 U 1 HE ...

Page 5: ...quency input is 500 feet 150 meters 25 HP 25 TE 3 U 3 HE 4 13 16 122 5 5 1 16 128 5 4 x M2 5x11 1 U 1 HE 1 750 inches 44 5 mm 1 HP 1 TE 0 200 inch 5 1 mm 19 inch 486 2 mm rack conforms to DIN 41494 Part 5 and IEC 297 3 Not included with Model 3300 3500 5 6 142 2 28 HP 28 TE Optional screw terminal Rear rail for mounting connectors that conform to DIN 41612 and IEC 603 2 Not included with Model 330...

Page 6: ...only and a plug in connector for power supply wiring Working from the front of the rack use the supplied M2 5x8 screws to install the wiring connectors onto the back of the rack as follows 1 Use the centers of the guide rails as reference points and refer to Figure 3 page 6 2 Install the input output wiring connector at 4 HP 4 TE from the neighboring unit or from the edge of the rack 3 Model 3500 ...

Page 7: ...the guide rails Power supply wiring connector is 25 HP 25 TE from neighboring unit or from edge of rack Input output wiring connector is 4 HP 4 TE from neighboring unit or from edge of rack Keyed sensor wiring connector is 16 HP 16 TE from neighboring unit or from edge of rack Keys on sensor wiring connector M2 5x8 Install screws and connectors from front of rack Model 3500 has six M2 5x8 screws a...

Page 8: ...ut wiring terminal card STEP 5 Connecting the Model 3500 to the sensor To connect the Model 3500 transmitter to a Micro Motion sensor follow the steps below If you are installing the Model 3300 applications peripheral this step is not required Table 1 Input output wiring terminals Terminal number Designation c 2 a 2 Primary 4 20 mA output c 4 a 4 Secondary 4 20 mA output c 6 a 6 Frequency input c ...

Page 9: ... instructions 4 At the sensor a Clip the cable drain wires b Connect wiring inside the junction box housing and tighten the screws to hold the wires in place For information on your sensor s junction box terminals see the sensor installation manual or Micro Motion s 9 Wire Flowmeter Cable Preparation and Installation Guide 9 wire cable from sensor Model 3500 Keyed sensor wiring connector see Figur...

Page 10: ...e power supply ground directly to earth ground Keep all ground leads as short as possible Ensure that all ground wiring has less than 1 ohm impedance 3 Connect wires to the upper and lower terminals A user supplied switch may be installed in the power supply line For compliance with low voltage directive 73 23 EEC European installations a switch in close proximity to the rack is required CAUTION I...

Page 11: ...10 Figure 6 Power supply wiring terminals Power supply wiring terminals Terminal designations for AC power Terminal designations for DC power ...

Page 12: ...pan Emerson Process Management Shinagawa NF Bldg 5F 1 2 5 Higashi Shinagawa Shinagawa ku Tokyo 140 0002 Japan T 81 3 5769 6803 F 81 3 5769 6843 Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T 65 6777 8211 F 65 6770 8003 Micro Motion Inc USA Worldwide Headquarters 7070 Winchester Circle Boulder Colorado 80301 T 303 530 8400 800 522 6277 F 303 ...

Reviews: