14
RFT9709 Transmitter Instruction Manual
Wiring
continued
Figure 2-6. Wiring to ELITE
®
CMF sensors
0
1
2
3
4
5
6
7
8
9
Green
White
Brown
Violet
Yellow
Orange
Blue
Gray
Red
Brown
Red
Clip drain wire back
Green
White
Clip drain wire back
Blue
Gray
Clip drain wire back
Orange
Violet
Yellow
Clip drain wire back
ELITE
®
sensor
terminals
Flowmeter
cable
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Yellow
Black
(Drains from all
wire sets)
Maximum cable length 1000 ft. (300 m)
Prepare cable in accordance with the instructions
that are shipped with the cable
RFT9709
terminals
Optional remote temperature detector (RTD) wiring (required for API)
ELITE
®
sensor
terminals
No connection
1. In the sensor junction box, identify the violet, yellow, and
orange wires that come from the sensor.
2. Disconnect these wires from the terminal block, then tie
all three together with a wire nut to protect exposed wire
ends.
3. Connect three wires from the external RTD to the
terminals from Step 2, as illustrated at left. If the RTD has
a fourth wire, it remains unconnected.
Violet
Yellow
Orange
Blue
Gray
Red
Green
White
Brown
3-wire or 4-wire
RTD
Black (drains)
Brown
Red
Orange
Yellow
Green
Blue
Violet
Gray
White
09wire.fm Page 14 Monday, March 1, 1999 4:31 PM
Summary of Contents for Fisher-Rosemount RFT9709
Page 1: ...Model RFT9709 Transmitter Instruction Manual March 1999 ...
Page 2: ......
Page 4: ......
Page 8: ...iv RFT9709 Transmitter Instruction Manual ...
Page 34: ...26 RFT9709 Transmitter Instruction Manual ...
Page 52: ...44 RFT9709 Transmitter Instruction Manual ...
Page 60: ...52 RFT9709 Transmitter Instruction Manual ...
Page 62: ...54 RFT9709 Transmitter Instruction Manual ...
Page 72: ...64 RFT9709 Transmitter Instruction Manual ...
Page 73: ......