30
RFT9709 Transmitter Instruction Manual
Startup
continued
3.4
Totalizer reset
Mass and volume totalizers cannot be reset independently. When one
totalizer is reset, the other is also reset.
A special version of the ProLink program is required for totalizer control.
The HART Communicator cannot be used for totalizer control. (Contact
the factory for additional ProLink information.)
3.5
Process measurement
After flowmeter zeroing has been completed as described in
Section 3.2, page 27, the flowmeter is ready for process measurement.
WARNING
When the totalizers are stopped, the frequency/pulse
output is disabled.
If the frequency/pulse output is used for process control,
failure to set control devices for manual operation could
affect process control.
• Before stopping the totalizers, set process control
devices for manual operation.
• To enable the frequency/pulse output, restart the
totalizers.