PRECISE-LOG Data Loggers User’s Manual
Page 40/57
Sampling Interval
1 second to 9 seconds in 1-second increment
10 seconds to 50 seconds in 10-second increment
1 minute to 59 minutes with 1-minute increment
1 hour to 12 hour with 1-hour increment
Making changes to the Sampling Interval will affect Total Time Span fields.
On-Board LED
Check this field to enable the on-board status LED. If the LED is enabled it will flash each
time when it samples data to indicate:
1.
The logging is active if the LED flashes in green.
2.
The logger is in alarm state if the LED flashes in red.
3.
The battery will die soon if the LED flashes in amber.
If you do not need LED indication, you can uncheck this field to increase the battery life.
When Memory Full
If you want the logger to stop logging when the memory is full select
Stop logging
. If you
want the logger to continue logging and overwrite the oldest data with the new data (FIFO),
you choose
Continue logging
.
Time to Start & Time to End
These two fields specify the desired time the logger will start logging data and the time to
stop logging data.
Making changes to the Time to Start/End will affect Total Time Span fields mentioned later.
If you have selected the
Continue logging
field, when the memory is full, both start time and
end time will move forward accordingly.
Total Time Span
These fields are an alternate way to specify the total logging time from the start time you
specified above.
Changes made on these fields will affect
Time to End
field.
Channel Settings:
Enabled
Check this field to enable this channel for logging.