Guidance and manufacturer’s declaration-electromagnetic immunity
The BP3GT1-6B is intended for use in the electromagnetic environment specified below.
The customer or the user of the BP3GT1-6B should assure that is used in such and environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the BP3GT1-6B including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d = 1,2
√
P
d = 1,2
√
P
80MHz to 800 MHz
d = 2,3
√
P
800MHz to 2,5 GHz
Where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the BP3GT1-6B is used exceeds the applicable RF
compliance level above, the BP3GT1-6B should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the BP3GT1-6B.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Summary of Contents for BP3GT1-6B
Page 1: ...Microlife BP3GT1 6B Instruction Manual v01 1 Model BP3GT1 6B ...
Page 2: ...Microlife BP3GT1 6B Instruction Manual v01 2 ...
Page 3: ...Microlife BP3GT1 6B Instruction Manual v01 3 ...
Page 4: ...Microlife BP3GT1 6B Instruction Manual v01 4 ...
Page 5: ...Microlife BP3GT1 6B Instruction Manual v01 5 ...
Page 6: ...Microlife BP3GT1 6B Instruction Manual v01 6 ...
Page 7: ...Microlife BP3GT1 6B Instruction Manual v01 7 ...
Page 8: ...Microlife BP3GT1 6B Instruction Manual v01 8 ...
Page 9: ...Microlife BP3GT1 6B Instruction Manual v01 9 ...
Page 10: ...Microlife BP3GT1 6B Instruction Manual v01 10 ...
Page 11: ...Microlife BP3GT1 6B Instruction Manual v01 11 ...
Page 12: ...Microlife BP3GT1 6B Instruction Manual v01 12 ...
Page 13: ...Microlife BP3GT1 6B Instruction Manual v01 13 ...
Page 14: ...Microlife BP3GT1 6B Instruction Manual v01 14 ...
Page 15: ...Microlife BP3GT1 6B Instruction Manual v01 15 ...
Page 16: ...Microlife BP3GT1 6B Instruction Manual v01 16 ...
Page 17: ...Microlife BP3GT1 6B Instruction Manual v01 17 ...
Page 18: ...Microlife BP3GT1 6B Instruction Manual v01 18 ...
Page 19: ...Microlife BP3GT1 6B Instruction Manual v01 19 ...
Page 20: ...Microlife BP3GT1 6B Instruction Manual v01 20 ...
Page 21: ...Microlife BP3GT1 6B Instruction Manual v01 21 ...
Page 22: ...Microlife BP3GT1 6B Instruction Manual v01 22 ...
Page 23: ...Microlife BP3GT1 6B Instruction Manual v01 23 ...
Page 24: ...Microlife BP3GT1 6B Instruction Manual v01 24 ...
Page 25: ...Microlife BP3GT1 6B Instruction Manual v01 25 ...
Page 28: ...Microlife BP3GT1 6B Instruction Manual v01 28 ...