background image

GlucoTeq - BGM 200

IB BGM 200 EN 2620 - Revision Date: 2020-07-02

Explanation of Symbols

For 

In vitro

 diagnostic use

Temperature limitation

Please consult instructions for use

Use by / Expiry date

Do not reuse

Manufacturer

Lot number

Caution, consult accompanying document

Keep dry

Keep away from sunlight

Humidity limitation

EU representative

This product fulfils the requirements of Directive 98/79/EC in vitro diagnostic 

medical device.

Do not use if package is damaged

Serial number

Distributor

Microlife AG

Espenstrasse 139

9443 Widnau

Switzerland

www.microlife.com

Wholecare 

Biomedical Corporation

8F, 443, RuiGang Road, NeiHu

Taipei, 11492, Taiwan

Medical Device Safety Service GmbH

(MDSS GmbH)

Schiffgraben 41

30175 Hannover / Germany

LOT

0123

0123

Overview

1.Before you begin

2.Getting to know your system

3.Prepare for blood sampling

4.Performing blood test

5.Alternative site testing

6.Control solution testing

7.Memory recall

8.Caring for your meter and test strip

9.Specifications

10.System Troubleshooting

11.Performance Characteristics

12.Guarantee

1. Before you begin

Please read this before using.

The following basic safety precautions should always be taken.

1.Close supervision is necessary when the device is used by, on, or near children or people with disabil-

ities.

2.Use the device only for the intended use described in this user guide.

3.Do not use test strips and control solutions which are not supplied by the manufacturer.

4.Do not use the device if it is not working properly, or if it has suffered any damage.

5.Before using any product to test your blood glucose, read all instructions thoroughly and practice the 

test. Do all quality control checks as directed and consult with a diabetes healthcare professional.

Keep this user guide with you.

Intended use:

Principle of Measurement

Blood glucose is measured by an electrical current that is produced when a blood sample mixes with 

the reagent (special chemicals) of the test strip. The electrical current changes with the amount of 

glucose in the blood sample. The Microlife GlucoTeq meter measures the strength of the electrical 

current, calculates your blood glucose level and then displays your result in either milligrams of glucose 

per deciliter (mg/dL) or millimoles of glucose per liter (mmol/L).

Caution

1.The user should not take any decision of medical relevance without first consulting his or her medical 

practitioner.

2.Call your doctor immediately if you experience symptoms that are not consistent with your blood 

glucose test results.

3.High altitudes above 3,402 meter (11,161 ft) may affect the test results.

4.Temperatures outside the range of 10°C to 40°C (50°F to 104°F) may affect the test results. Do not 

test outside of these temperature ranges.

5.Do not perform servicing and maintenance while the meter is in use.

6.Modification of this equipment is not allowed. If the equipment is used in a manner not specified by 

the manufacturer, the protection provided by the equipment may be impaired.

7.Do not use this meter near cellular or cordless telephones in a call, walkie-talkies, garage door 

openers, radio transmitters, or other electrical or electronic equipment that are sources of electromag-

netic radiation, as these may interfere with the proper operation of the meter.

Important Health-Related Information

1.Apply only capillary whole blood sample to test your blood glucose. Applying other substances or 

plasma, serum will cause wrong results.

2.Severe dehydration and excessive water loss may cause false low results. If you believe you are 

suffering from severe dehydration, consult your healthcare professional immediately.

3.Test results below 60 mg/dL (3.3 mmol/L)*

1

 indicates low blood glucose (hypoglycemia). Test results 

greater than 240 mg/dL (13.3 mmol/L)*

2

 indicates high blood glucose (hyperglycemia).If your results 

are below 60 mg/dL (3.3 mmol/L) or above 240 mg/dL (13.3 mmol/L), repeat the test, and if the results 

are still below 60 mg/dL (3.3 mmol/L) or above 240 mg/dL (13.3 mmol/L), consult your healthcare 

professional immediately.

4.Inaccurate results may occur in severely hypotensive individuals or patients in shock. Inaccurate low 

results may occur for individuals experiencing a hyperglycemic-hyperosmolar state, with or without 

ketosis. Critically ill patients should not be tested with blood glucose meters.

5.Abnormal red blood cell counts may cause false results. Please consult your healthcare professional 

if you do not know your hematocrit level.

6.Interference: Reducing substances occurring in the blood naturally (uric acid, bilirubin) or from thera-

peutic treatments (ascorbic acid, acetaminophen) will not significantly affect Microlife GlucoTeq test 

results. However, elevated concentrations of these substances may affect test results. The 

compounds listed in the table were found to have no affect at the concentration indicated.

REFERENCE:

Kahn, R. and Weir, G.: Joslinis Diabetes Mellitus, 13thed Philadelphia: Lea and Febiger (1994), 489. *1

Krall, L.P. and Beaser, R. S.: Joslin Diabetes Manual. Philadelphia: Lea and Febiger(1989), 261-263. *2

2. Getting to know your System
The Microlife GlucoTeq Glucose Monitoring System

The Microlife GlucoTeq system uses the latest technology to provide you with easy and comfortable 

testing. The system requires only 0.5uL blood sample to complete the testing in only 5 seconds.

Please review the contents of your purchase to confirm that all the components are included as listed 

on the side of your meter box.

The system is intended for use outside the body (

in vitro

 diagnostic use only). It should be used 

only for self-testing blood glucose (blood sugar) and only with fresh capillary whole blood 

samples. The system is intended for use in the home and in clinical settings. The system should 

not be used for the diagnosis of diabetes or for the testing of newborns.

Compounds

Concentrations higher than the following values may cause 

inaccurate results

Acetaminophen

8.0 mg/dL (0.53 mmol/L)

Ascorbic Acid

5.0 mg/dL (0.28 mmol/L)

Aspirin

60 mg/dL (3.33 mmol/L)

Bilirubin

90 mg/dL (1.54 mmol/L)

Cholesterol

500 mg/dL (12.9 mmol/L)

Creatinine

5.0 mg/dL (0.44 mmol/L)

Dopamine

2.0 mg/dL (0.11 mmol/L)

EDTA

360 mg/dL (12.3 mmol/L)

Galactose

900 mg/dL (50 mmol/L)

Gentisic Acid

5.0 mg/dL (0.32 mmol/L)

Glutathione

53 mg/dL (1.72 mmol/L)

Haemoglobin

500 mg/dL (0.08 mmol/L)

Heparin

8,000 U/dL

Hydroxyurea

3.0 mg/dL (0.39 mmol/L)

Ibuprofen

50 mg/dL (2.42 mmol/L)

Icodextrin

13 mg/dL (0.01 mmol/L)

L-dopa

10 mg/dL (0.51 mmol/L)

Maltose

900 mg/dL (26.3 mmol/L)

Methyldopa

3.0 mg/dL (0.13 mmol/L)

Pralidoxime Iodide

25 mg/dL (0.94 mmol/L)

Salicylate

60 mg/dL (4.34 mmol/L)

Tolazamide

100 mg/dL (3.21 mmol/L)

Tolbutamide

400 mg/dL (14.8 mmol/L)

Triglycerides

2,000 mg/dL (22.6 mmol/L)

Uric Acid

8.0 mg/dL (0.48 mmol/L)

Xylose

100 mg/dL (6.66 mmol/L)

1

Illuminated Test Strip Slot

Insert the test strip here. The meter will turn on auto-

matically and the slot LED light will illuminate.

To track the position of test strip slot without strip 

inserting, press main button

and the slot LED light will be flashing.

2

LCD Display

Guides you through the test using symbols and 

simple messages.

3

Main Button

Turns the meter on or performs other functions 

described in this user guide.

4

Data Port

Allows you to transfer the information stored in the 

meter to a computer to view, analyze and print.

5

Set Button

Located on the back of the meter, inside the battery 

compartment; used to set up the meter.

6

Battery Compartment

Holds ONE 3V Lithium battery (battery type 

CR2032). Please install battery into meter before 

you start testing.

7

Meter Label

Each meter has its unique number on it. Do not alter 

or tear the label off.

vcc

GND

1

2

3

3

4

5

6

7

If any part of the display is not working, contact your local distributor for help.

Battery

Microlife GlucoTeq meters come with one (1) CR2032 lithium battery. The battery is manufactured in a 

charged state and not designed for recharging. Recharging the battery can cause battery leakage, or 

in some cases, high pressure rupture.

Warning:

 Batteries can explode or leak and cause burns if installed backwards, disassembled, 

charged or exposed to water, fire or high temperatures.

Low Battery

The meter will alert you when the power is getting low by displaying two different messages:

When the battery symbol 

AR

 appears alone on the display, the meter is functional and the results 

remain accurate, but you should change the battery as soon as possible.

When the battery symbol 

AR

 appears together with the 

«E-b»

 symbol on the display, the battery does 

not have enough power for a test. You must change the battery before using the meter.

Replacing the battery does not affect the meter’s memory (previous test results stored in 

memory). However, the date, time and unit settings may need to be updated; update the settings 

by following the steps in «Setting Time and Date». 

As with all small objects, the battery should be kept away from small children. If the battery is swal-

lowed, seek medical assistance immediately.
Remove the battery if you are not going to use the device for a long period of time (i.e. 3 months 

or more), or the battery might leak chemicals.

Please discard the product or the batteries properly according to the regulations of your country.

Setting Mode

Please install battery first and complete the setting before you begin to test.

8

Contact Bars

Insert this end of the test strip into the meter. Push 

it in firmly until it will not go any further.

9

Test Strip Handle

Hold this part to insert the test strip into the test strip 

slot on the meter.

AT

Confirmation Window

Shows whether enough blood has been drawn into 

the test strip’s absorbent channel.

AK

Absorbent Channel

Apply a drop of blood and it will be drawn in auto-

matically.

AL

Test Result Area

Displays test results.

AS

Unit of Measure

Indicates what unit of measure 

the test result is displayed in.

AM

Appears when test result is 

lower than 70 mg/dL (3.9 

mmol/L) or higher than 180 

mg/dL (10 mmol/L).

BT

Temperature Symbol

Appears when temperature is 

outside of operating range.

Appears when test result is 

within the range of 70 to 120 

mg/dL (3.9 mmol/L to 6.7 

mmol/L).

BK

Control Solution Test 

Symbol

Shows that you are in control 

solution mode.

AN

Ketone Symbol

Appears when test result is 

higher than 240 mg/dL (13.32 

mmol/L).

BL

Memory Symbol

Appears when you review the 

memory.

AO

Alarm Symbol

Appears when you are 

setting alarms.

BM

After Meal

Indicates that your test is an 

After Meal test.

AP

Test Strip Symbol

Appears when the meter is in 

testing mode.

BN

Before Meal

Indicates that your test is a 

Before Meal test.

AQ

Blood Drop Symbol

Flashes when sample should 

be applied.

BO

Date

AR

Low Battery Symbol

Appears when the battery 

power is low.

BP

Time

STEP 1: 

With the meter off, press 

the   mark on the battery cover 

and slide up to open the cover.

STEP 3:

 Replace with a new 

battery. Be sure to align the 

battery properly with the positive 

(+) side up.

STEP 2:

 Remove the old battery 

by pressing on the buckle beside 

the battery and gently pulling it at 

the same time.

STEP 4:

 Close the battery cover. 

If the meter does not power on 

and enter setting mode after you 

have replaced the battery, check 

that the battery is correctly 

installed.

STEP 1:

 Enter Setting Mode

If your meter is off, press the 

SET

 

button 

5

 located in the battery 

compartment. The meter is now 

in the setting mode.

STEP 4:

 Set the Date

Press and release the main button 

3

 to advance the date. With the 

correct date on the display, press 

the 

SET

 button 

5

 and the time 

will appear on the display with the 

hour segment flashing.

STEP 2:

 Set the Year

Press and release the main button 

3

 to advance the year. With the 

correct year on the display, press 

the 

SET

 button 

5

 and the date 

will appear on the display with the 

month segment flashing.

STEP 5: 

Set the Hour

Press and release the main 

button 

3

 to advance the hour. 

With the correct hour on the 

display, press the

 SET

 button 

5

 

and the minutes segment will 

start flashing.

STEP 3:

 Set the Month

Press and release the main button 

3

 to advance the month. With the 

correct month on the display, 

press the 

SET

 button 

5

 and the 

date segment will start flashing.

STEP 6:

 Set the Minutes

Press and release the main 

button

3

 to advance the minutes. 

With the correct minutes on the 

display, press the 

SET

 button 

5

 

and the 

«OFF»

 will start flashing.

Fig. A

STEP 7: 

Set the Alarm

The meter allows you to set 4 different alarms with an order from 1 to 4. 

 

will be displayed during the alarm setting.

Fig. B 1 - Press the main button

3

 to turn the alarm ON (Fig.A) or OFF (Fig.B).

2 - Press 

SET

 button 

5

 to move to hour setting for alarm. (Fig.C) Use the 

main button 

3

 to set your desired hour. 

Press 

SET

 button

5

 again to set the minute. (Fig.D).

Fig. C 3 - Press 

SET

 button to set the next alarm.

4 - Repeat step 1-3 to set the second, third and fourth alarm. 

After completing fourth alarm, the AC/PC will start flashing.

Fig. D

STEP 8:

 Set AC/PC

Press and release the main button 

3

 to select ON/OFF. With the correct 

setting, press the 

SET

 button 

5

 for confirmation.

STEP 9: 

Delete Memory

If you do not want to clear the memory, press the 

SET

 button again to skip 

this step.

If you want to clear ALL memory, press the main button for five (5) seconds. 

The 

«- - -»

 will appear to indicate all memory has been deleted and the meter 

will turn off.

STEP 10:

 Complete Setting

Setup is now complete. 

«OFF»

 is displayed and the meter will turn off.

8

9

AT

AK

AL

AM
AN

AO AP AQ

AR

AS

BT

BK

BL

BM

BN

BP

BO

vcc

GND

vcc

GND

vcc

GND

vcc

GND

3. Prepare for blood sampling

If your lancing device did not come with a lancing device insert, the following information shall take 

the place of the insert.

Microlife LD-100 Lancing Device

Your lancing device and lancets are used for obtaining blood samples from the puncture site.

Special Features

Intended Use: The lancing device is intended for use by a single person with the lancet puncturing 

through the skin for collecting blood sample.

Safety Switch design: when the depth adjustable cap is open, the trigger will be relieved automatically 

which prevents users from being pricked if you are not intended to start the test. You can eject the used 

disposable lancet on one hand operation by pushing the plunger with your thumb at this status.

When the cap is closed, it will lock the lancet eject function and prevent the lancet from being removed 

when you are intended to start the test.

Stable and sophisticated structure design ensures almost no vibration, so that the lancet pricking is 

steadier with almost no pain. It lowers the psychological pressure of users very much and let people 

recover the test frequencies which recommended by HCPs.

A stylish, slim and ergonomic design is very convenient and user friendly.

Clear indication for depth penetration (numeric display): easy to read and adjust for users.

Important Lancing Device and Lancets Information

1.Lancet is for single use only.

2.Keep lancing device and lancets clean.

3.Use caution when removing the used lancet from the device and when disposing the used lancet.

The meter and lancing device are for single patient use.
Do 

NOT

 share them with anyone including family members.

Do 

NOT

 use on multiple patients.

Setting your Lancing Device

4. Performing a Blood Test

1

A stylish, slim and ergonomic design is very 

convenient and user friendly.

2

Big window to show the penetration depth with 

numeric display which is easy for the user to 

read.

3

Big release button, comfortable and easy to 

operate.

4

Patented feature of lancet ejection by one 

thumb pushing with only one hand operation.

5

Patented safety switch design:

when the cap is closed, it allows only load but 

lock the function of lancet ejection. When cap 

is open, it allows lancet ejection but release the 

trigger automatically to prevent accidental 

pricking.

6

Depth adjustable cap

7

Release button

8

Plunger

9

Hub

AT

Clear cap (optional)

AK

Disposable lancet (optional)

1. Hold the depth adjustable 

cap in one hand and hold 

hub in the other hand, twist 

off the depth adjustable 

cap.

5. Select a depth of pene-

tration by rotating the top 

portion of the depth adjust-

able cap until the number 

matched to the window.

Settings are based on skin 

type.«1,2» are for

soft or thin skin. «3,4» are 

for average skin.

«5,6» are for thick or 

calloused skin.

2. Insert a new disposable 

lancet firmly into the lancet 

holder.

6. Hold the hub in one hand 

and push on the plunger in 

the other hand. The device 

will be cocked.

Release the plunger; it will 

go back to its original posi-

tion near the hub.

3. Twist off and set aside 

the protective cover of the 

disposable lancet.

4. Replace the depth 

adjustable cap.

The Lancing device is made by
MEDIFUN CORPORATION

No.8, Shuyi Rd., South Dist., Taichung 

City, 40241, Taiwan (R.O.C.)

1. Wash Your Hands and the Puncture Site:

Wash your hands in warm, soapy water. Rinse and dry completely. Warm your 

fingers to increase blood flow.

2. Insert Test Strip:

 

Remove a new test strip from vial. Be sure to tightly replace vial cap after 

removing test strips. Insert a test strip with the contact bar end entering into the 

test strip slot first. Push the test strip as far as it will go without bending it. The 

meter turns on automatically.

*To track the position of test strip slot, press main button and the slot LED light will flashing.

3. Select and Lance a Puncture Site:

Hold the prepared lancing device firmly against the side of your fingertip. Press 

the release button.

NOTE:

 If you want to do alternative site testing, please refer to the «About Alter-

native Site Testing (AST)» section. Please consult your healthcare professional 

before obtaining blood from site other than your fingertip.

4. Obtain a Blood Sample:

Gently massage your finger or puncture site to obtain the required blood 

volume. To perform the test, you need only 0.5 

μ

L of blood sample. Do not 

smear the blood sample. To obtain best accurate result, wipe off the first drop 

of blood and gently squeeze another drop of blood.

5. Apply Blood Sample:

When the meter shows the «

» symbol, apply blood to the opening of the absor-

bent channel of the test strip where it meets the narrow channel. Blood will be 

drawn into the test strip. If the test strip confirmation window is full, you will hear 

a beep.

6. Read Your Result:

After the meter counts down from 5 to 1, your blood glucose test result appears 

along with the unit of measure, date and time.

7. Turn Off the Meter:

This blood glucose result is automatically stored in the meter memory. Turn the 

meter off by removing the test strip. Discard the used test strip carefully to avoid 

contamination.

8. Remove the Adjustable Comfort Tip:

After use, put the Protective Cap back onto the exposed needle of the lancet.

9. Dispose the used lancet correctly:

Slide the Lancet Ejector forward and dispose the used lancet in an approved 

container. Dispose the used lancet according to your country's safety regula-

tions. Replace the Lancing Device Cap.

1

2

3

4

5

6

7

9

8

AT

AK

1

5

2

6

3

4

AM

PM

Incorrect

Correct

Model No: BGM 200

Gluco

Teq

Blood Glucose Monitoring System

EN

Instruction Manual

Reviews: