Sample Preparation
VacPrep 061
2-14
Feb 2013
Degassing Samples
The VacPrep 061 offers three methods of sample preparation:
•
Vacuum preparation
- The sample is heated as a vacuum is applied to the sample tube
to remove moisture and other contaminants. This method should be used only with
materials which are not likely to fluidize.
•
Vacuum-backfill preparation
- The sample is heated as the operator alternately applies
a vacuum and then flows gas into the sample tube to remove moisture and other
contaminants. The alternating process may be repeated several times. The VacPrep 061
is designed for low-pressure backfilling only.
•
Flowing gas preparation
- The sample is heated as the operator flows gas into the
sample tube to remove moisture and other contaminants.
The following topics describe the procedures to degas samples using each method.
Do not exceed 5 psi regulator pressure when backfilling sample tubes. If the gas
pressure is greater than the recommended maximum of 5 psig (35 kPag), the sample
tube could be ejected from the fitting or broken.
Samples that contain a high level of moisture could benefit from a higher flow rate
than 5 psi regulator pressure when using the flowing gas method. A flow rate up to 50
cm
3
/min (approximately 16 psi regulator pressure) may be used to aid in purging
moisture from the sample when using flowing gas preparation only.
Summary of Contents for 061-00023-00
Page 1: ...VacPrep 061 Operator s Manual 061 42802 01 Feb 2013 ...
Page 2: ... Micromeritics Instrument Corporation 2002 2013 All rights reserved ...
Page 4: ......
Page 46: ...Sample Preparation VacPrep 061 2 26 Feb 2013 ...
Page 56: ......
Page 60: ...Ordering Informatioin VacPrep 061 4 4 Feb 2013 ...