VacPrep 061
Sample Preparation
Feb 2013
2-23
7. Using the appropriate heating station adapter, place the sample tube in a vacant heating
station.
8. Allow the sample to degas for the desired length of time. One to two hours is typical for
many samples at temperatures above 200 ºC. To determine the optimum time, select the
highest safe temperature for your sample, then degas it for various intervals, determining
the surface area after each interval.
For example, degas the sample for 15 minutes, then measure the surface area. Repeat this
process until the surface area no longer increases. The length of time required to reach this
point is the minimum amount of time required to degas the sample.
9. After the sample is sufficiently degassed, remove the sample tube from the heating station.
DO NOT TURN OFF THE GAS FLOW KNOB.
Keep the gas delivery tube and stopper in the sample tube while cooling to prevent
contamination from the ambient atmosphere.