Maintenance
Checking Gas Flow
The gas flow through the flexible tubing and gas delivery tubes should be checked once a
week to ensure that there are no obstructions or leaks in the tubing. To check the flow, place
the flow control valve in the GAS position. Insert the gas delivery tube in a beaker of water.
Bubbles should appear. Approximately the same number of bubbles should appear in the bea-
ker of water when moved to each station.
If bubbles do not appear for a particular station, detach the suspect gas delivery tube from the
flexible tubing and insert the flexible tubing in the beaker of water. If bubbles appear, the
problem is a clogged gas delivery tube. If the tube cannot be unclogged, replace it with a new
one (refer to
Cleaning and Replacing Gas Delivery Tubes,
). If bubbles do not
appear, check the valve inlet tubing (refer to
Connecting or Replacing Valve Inlet Tubing,
).
If gas flow is not present in other stations, refer to the following table.
Table 3-2. Gas Flow Problems
Gas delivery tubes and flexible tubing may be ordered in sets of six from
Micromeritics. Refer to Chapter 4, Ordering Information.
Why
What To Do
Gas pressure is set too low.
Increase the flow rate by turning the regulator pressure
control knob clockwise until the correct pressure gauge
reading is shown. (Refer to
Gas bottle is empty.
Replace the gas bottle.
Valve inlet tubing is
disconnected.
Check valve inlet tube connections, refer to
.
Flow control valve is
damaged.
Call the appropriate service personnel.
Summary of Contents for 061-00023-00
Page 1: ...VacPrep 061 Operator s Manual 061 42802 01 Feb 2013 ...
Page 2: ... Micromeritics Instrument Corporation 2002 2013 All rights reserved ...
Page 4: ......
Page 46: ...Sample Preparation VacPrep 061 2 26 Feb 2013 ...
Page 56: ......
Page 60: ...Ordering Informatioin VacPrep 061 4 4 Feb 2013 ...