AquaPrep II 056
Installing the AquaPrep
Aug 2011
2-5
Installing the Tubing
The AquaPrep has inlet and outlet connectors onto which tubing is installed for circulating the
water through the AquaPrep for degassing.
Install the two pieces of tubing and hose clamps (supplied in the accessories kit) onto the con-
nectors; one located on the left side of the front panel and the other on the end of the capsule.
Squeeze the clamps until 2 to 3 teeth are engaged.
Connecting the Power Cord
Insert one end of the power cord into the input power connector on the rear panel of the Aqua-
Prep and the other end into an appropriate power source.
Do not turn the AquaPrep on until you you have prepared the water container and are ready to
begin operation. Refer to Chapter 3 for complete operating instructions.
Tubing Connectors
Do not turn the AquaPrep on until the water is prepared and ready for deaeration.
Failure to have water circulating while the unit is turned on may damage the unit.