DVVA II 4000
Sample Analysis
400-42803-01 (Rev B) - July 2017
6-3
Sample Analysis
Use this command to perform an analysis. A blank analysis is part of the analysis procedure and
is performed before you add the sample material for analysis. The measured force includes
changes in friction caused by the sample type
A blank analysis is performed to correct for this friction. You can skip the blank analysis if you
check the
Use Prior Blank
option; the data from the last blank analysis will be used.
Choose a file from the Open Sample Information File dialog, or create a new one, to display the
Analysis dialog. The fields now contain the values for the selected file or, if creating a new file,
the specified defaults.
View
Allows you to view:
• the current operation
• the instrument schematic; refer to page
6-11
for a description
of the components on the schematic
• the instrument log; refer to page
6-14
for a description of the
instrument log