DVVA II 4000
Controls, Indicators, and Connectors
400-42803-01 (Rev B) - July 2017
2-3
Rear Panel
USB port
Power
connector
Voltage
selection
Circuit
breakers
Test points
Master
power
switch
USB port
For connecting the communications cable to the computer
Voltage selector switch
Enables you to select the correct operating voltage
Power connector
For connecting the power cord
Actuator and lower
piston circuit breakers
A trip of one of these breakers may indicate a motor failure.
Should this occur, wait a few minutes, reset the breaker, and retry
your operation. If the breaker trips again, contact your service
representative.
Test points
For service personnel only
Master power switch
Master power for the analyzer