DVVA II 4000
Troubleshooting
400-42803-01 (Rev B) - July 2017
9-11
If, after several attempts performing steps 4 and 5, the piston cannot be moved to its home
position:
a. Close the safety shield.
b. Press the
Reset
button.
c. Proceed to
“Action 2” on page 9-12.
7. Select
Unit 1 > Clean System f
rom the main menu to clear the compressed pellet from the
compression chamber.
8. The DVVA II should now be operational.