Troubleshooting
DVVA II 4000
9-12
400-42803-01 (Rev B) - July 2017
Action 2
If Action 1 does not solve the problem, you can manually release the piston as described below.
1. Support the lower front panel from the bottom and remove the four screws shown in the
photograph below.
2. Remove the panel and set it aside.
Never override the instrument panel sensor or operate the instrument with the panel
removed.