Maintenance
DVVA II 4000
9-16
400-42803-01 (Rev B) - July 2017
Replacing and Conditioning the Piston Tip
The piston tip should be replaced when deemed necessary; and must be conditioned before
using.
Replacing the Upper Piston Tip
1. Select
Unit 1 > Return Upper Piston Home
to move the piston to its uppermost position,
completely out of the sample chamber.
2. Slide the sample compartment shield to its uppermost position.
3. Place a stiff piece of paper or cardboard over the top of the sample funnel to prevent the
loosened tip from accidentally falling down into the chamber.
Upper piston
home position