Maintenance
DVVA II 4000
9-18
400-42803-01 (Rev B) - July 2017
Replacing the Lower Piston Tip
1. Remove the lower piston tray.
2. Using a 3/8-in. wrench, rotate to the left to loosen and remove the piston in the sample tray.
3. Install a new piston tip. Using the same wrench, moderately tighten the piston tip.
4. Replace the lower piston tray and press the
Reset
button on the analyzer.
5. Select
Unit 1 > Unit configuration
to display the Unit configuration dialog.
6. Click the
Reset
buttons to reset the
Lower Tip Run count
to 0 (zero), then click
OK
to close
the dialog.