Maintenance
DVVA II 4000
9-24
400-42803-01 (Rev B) - July 2017
26. Enter the correct mass for the sample being analyzed.
27. Repeat steps 13 through 24 for the remaining nine samples. Be sure to:
• change the Sample identification sequentially
• wipe the piston tip after each blank analysis and sample analysis
28. Advance to
Overlaying Results
to determine if the piston is sufficiently conditioned.
Overlaying Results
When results start to be consistent, the piston tip is sufficiently conditioned. Perform the follow-
ing steps to overlay the results from the last six analyses:
1. Select
File > Open Sample Information
and choose the fifth file of the condition series.
2. Click the Report Options tab to display the Report Options dialog.