DVVA II 4000
Normalization
400-42803-01 (Rev B) - July 2017
9-31
Verifying Normalization Results
The current reference sample analyses are compared to the original "master" reference analysis
averages to determine by how much the current analysis has shifted from that average. The nor-
malization routine calculates a set of polynomials to correct for that difference. What is shown in
the Normalization window (see example below) is how well the software was able to determine
the polynomials it will use to correct, or normalize, the collected data.
1. Select
Unit > Normalization
; the Normalization dialog is displayed.
2. Note the residual values displayed on the dialog.