DVVA II 4000
Appendix A
400-42803-01 (Rev B) - July 2017
A-1
A. ERROR MESSAGES
Error messages for the DVVA II program are listed numerically. If the
Action
response instructs
you to contact your service representative, record the error message and make backup copies of
any files involved in the operation.
2200 and 2400 Series
2201-
Cannot execute report subsystem.
2401-
FATAL ERROR: (error message).
2430-
Error accessing file (file name), error code = (number).
Cause:
Start Report failed to execute the report subsystem (which is a separate
process).
Action:
Restart the computer. If the problem persists, reinstall the application
(this will not affect any of your sample files). If the problem continues,
contact a Micromeritics service representative.
Cause:
An internal processing and/or hardware error has occurred.
Action:
Contact your service representative if you continue to receive this error
message.
Cause A:
A computer or network problem occurred.
Action A:
Check the performance of your computer devices or network.
Cause B:
A software error occurred when the file was accessed.
Action B:
Contact your service representative.