Appendix A
DVVA II 4000
A-12
400-42803-01 (Rev B) - July 2017
2511-
Print job (name) has been cancelled due to insufficient disk space. Delete unnec-
essary files and restart the report.
2512-
Print job (name) has been cancelled.
2513-
Unable to read the calibration file (file name).
2514-
Unable to write the calibration file (file name).
2515-
Warning: Changing the calibration information will affect the performance of the
instrument. Only qualified service personnel should do this. Do you wish to
proceed?
Cause:
The disk drive does not have enough space for the temporary file
required by the Windows Print Manager. Therefore, printing of the
requested report has been canceled.
Action:
Delete unnecessary files from the disk. You will require at least five
megabytes of free space for normal operation.
Cause:
The requested print job was canceled at your request.
Action:
None required.
Cause:
You selected an invalid calibration file or one that cannot be read.
Action:
Be sure the media containing the calibration file has no problems.
Cause:
An attempt to Save calibration data has failed due to possible media
problems.
Action A:
Be sure the media you want to Save the file to has no problems.
Action B:
Choose an alternate media to Save the calibration data.
Cause:
You have started the process of performing a calibration operation.
Action:
Calibration operations should only be done by, or under the direction
of, qualified service personnel.
Summary of Contents for DVVA II 4000
Page 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Page 4: ......
Page 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Page 16: ......
Page 32: ......
Page 116: ......
Page 136: ......
Page 160: ......
Page 180: ......
Page 212: ......
Page 214: ......
Page 240: ......
Page 246: ......
Page 248: ......